Quantima 技术参数
Part Number: 99-0552-00
RF发生器RF Generator
Free running, air cooled generator. Auto start from switch or keyboard.
Operating Frequency 40.68 MHZ
光栅Grating
1,800 L/mm t67holographic grating allows wide analysis range from 160 - 800 nm
and up to 6 pm resolution.
Optional 2400 l/mm grating allows analysis range from 160 - 640mnm with up to 4 pm resolution
Mass Flow Control
Nebuliser Gas flow regulated using mass flow controller.
Optional mass flow control available on Plasma and Auxilliary flows.
RF Power
Computer controlled from 500 W to 1,500 W with auto tuning.
Safety Interlocks
Interlocks on plasma compartment door, gas pressure, low/high mains voltage, plasma failure.
EMC Standards
Meets the following EMC standards:
EN655011, IEC801-2, IEC801-3 and IEC801-4.
Operating Voltage
200-240 V AC+ 10%, 3.5kVA, 20A
Frequency
50/60Hz
Dimensions
1520 mm (W) x 850 mm (H) x 730 mm (D)
Weight
245 kg
Accessories
99-0326-11 SDS 540 Autosampler
97-3764-00 Hydride Generator
40-0254-00 Programmable Temperature Spray Chamber
99-0322-00 Ultrasonic Nebuliser
99-0566-00 Ultrasonic Nebuliser with Desolvator
97-3749-00 Thermostating Module
97-3764-00 Humidifier
97-3718-00 Plasma Cam
97-3655-00 Inbuilt Chiller
97-3719-00 PIP
40-0182-00 2,400 l/mm High Resolution/Sensitivity Grating
97-3750-00 3 Channel Mass Flow Controller
97-2279-00 HF Sample Introduction Kit
95-0035-00 Organic Samples Kit
99-0562-00 Quantima Bench includes Storage Shelves and Wheels