High precision beam diagnostics for CW lasers
Patented technology: Unique tomographic image reconstruction of 2D/3D images
Versatile:Meaure beam profile, beam size, beam shape,position and power
Flexible: Wide spectral range, from 190nm through 1800nm
Accurate: Beam sizes from 3μm to 9mm with 0.1μm resolution
Compact:Based on USB 2.0 manifold box, measuring head, software
Main Features
New!12 bit A/D
High resolution sampling for all modes simultaneously
Real time beam profiles, beam size and gaussian fit
2D/3D plots of beam in real time
Beam centroid and ellipticity, Power measurement
Direct data logging to Excel files
Data streaming via RS232 or TCP/IP
Save image and snapshot files
Automatic Pass/Fail analysis report
Windows control library for user's application program
Analyzer USB
Patented Technology
The Beam Analyzer provides a bridging technology, producing the 3D intensity reconstruction of CCD cameras, while being capable of measuring very small spots at high resolution and huge dynamic range
The measuring technique is based on a multiple scanning knife-edge technology, combined with a tomographic image reconstruction for the creation of the 2D/3D display. When the drum spins, the knife-edges cut across the beam in an orthogonal plane to the direction of propagation. Astationary large detector inside the spinning drum measures light intensity. For attenuation, when needed, a built-in distortion free optical filter is inserted between the spinning drum and the detector. This technique provides the required attenuation without affecting beam quality. Each scanning knife-edge is oriented at a different angle on the drum and moves across the beam path in a different direction as the drum rotates.Consequently, during a single rotation of the drum, the instrument generates a set of profile curves, each representing the intensity profile of the beam from a different direction. This data is the input for the tomographic reconstruction algorithm to generate the 2D/3D intensity profile of the beam.
The Beam Analyzeris offered in two types of measuring heads:The BA7 uses seven individual knife-edges, providing more accurate measurement of the true beam shape and dimensions by gathering data from all 7 scans, while the BA3 uses only three knife-edges, and is recommended for smaller beams measurement as well as for a
near-Gaussian beams. The more knife-edges, the greater the level of detail obtained. For a beam distribution that is significantly non-Gaussian the BA7 would reconstruct a plot that closely matches the real beam.
System Presentations
Beam Analyzerprovides an extensive range of graphical presentations and analysis of laser beam parameters.
Beam Profiles and Width
Beam Analyzer’control software simultaneously displays two profiles from two
orthogonal knife-edges, or show each profile individually for greater visual detail.
These main profiles, located at 45 degrees from the base of the head, are called
“V” and “W”.
Beam widths are digitally displayed for any three user selected clip levels, with up to 0.1μm resolution. The beam profile displays are auto-scaled (optional) to maximize on-screen detail and resolution. Added detail can be obtained in a
special high-resolution mode. AGaussian fit profile can be overlaid on profiles in
real time, while the correlation and fit values are displayed digitally. Zoom
function automatically increases the displayed spatial resolution.
Chart mode
It is frequently necessary to monitor the Beam Width (or alternatively Beam Position) as a function of time. In Chart Mode, these parameters can be viewed in strip chart format, showing long term time-dependent stability or drift. The
measured data can be viewed on screen, saved or printed for further analysis.
System Presentations
2D and 3D Intensity Plots
The Projection function provides either a 2D or a 3D plot of the beam intensity profile, and is created using reconstructive tomography.The 2D contour maps and the 3D isometric plots can be displayed with or without scan axis and grids. These plots can be rotated along the beam optical axis, as well as be flipped. This feature enables the user to view the image from various angles around the beam. It is also possible to control the 3D plot wire density, as well as view the 3D
figure projections on XZ and YZ planes Data about the beam size and centroid position is displayed digitally along with centroid and beam size data.
PowerMeasurement
The beam power can be displayed either as a digital readout, or as a bar-type
display or in combination with an analog “needle”. Power presentation units
can be chosen as mW, μWor dBm. The user can offset the zero to deactivate the ambient light suppression. Pre-defined filter transmission files can be selected. Atest range can be defined and displayed to monitor beam power within specific limits, audio alarms are optional.
Beam Position and Ellipticity
The beam centroid position is continuously monitored relative to the center of the sensor area in real-time, along with beam shape, ellipticity (major and minor axes) and angular orientation. Azoom function is available for viewing the footprint of small beams more clearly. Observing beam position stability versus time is available upon setting up of the Chart function to monitor beam position at the
selected clip level.