性能参数
检测器 | 测量目标物 | 检测限 | |
光离子化检测器 (PID),;氢火焰离子化检测器 (FID);热导检测器 (TCD);硫磷检测器(FPD);远紫外 Far UV 吸收 (FUV) 检测器; 催化燃烧检测器 (CCD) | PID & FID: 有机物 (用PID 或FID测定VOCs ) 和无机物 (H2S, NH3, AsH3… ) ,可被PID灯离子化的化合物 (9.5, 10.6 或 11.7 eV) TCD: 通用检测器- fixed gases (O2, N2, CO…), 有机和无机化合物,对 H2, He, Xe 非常灵敏 FPD: 对含S和P的化合物选择性响应 Far UV: 对于所有的有机和无机化合物有近乎通用的响应 CCD: 测定可燃性化合物如烃类 | PID - 0.5 ppb 苯 FID - 50 ppb 苯 TCD - H2- 10 ppm; O2 150 ppm | |
动态线性范围 | 传感器 | 测量精度 | 响应时间 |
PID - 108 FID - 2 x 106 TCD - 2 x 104 FPD- S - 103; P 105 Far UV - 106 CCD - 5 x 104 | 1 或 2 个检测器 | ±1~2% 读出 | 依赖于保留时间 |
样品流速 | 进样 | 漂移 | 零点漂移 |
> 0.5 LPM | 柱上注射器进样,或自动六通阀气体进样(选项) | <1% ,24小时内 | NA- 每次运行自动归零 |
功耗 | 尺寸 | Weight: | 仪器操控条件 |
zei高< 50 瓦;115 VAC, 0.5 amps (12 VDC 5 amps) | Enclosure - 18.75''x 13.75'' x 7.5'' | Fiberglass case - 27.5 磅 | 5-40 ℃;0-95% RH (非冷凝) |
特点
PID Lamps | FID | FPD | 易于校正 |
Our 9.5 and 10.6 eV lamps remain the best in the business since we developed them over 30 years ago. | Our very sensitive FID has flame out sensor that will shut off the hydrogen if the flame goes out. | Our very sensitive FPD has flame out sensor that will shut off the hydrogen if the flame goes out. | 手动 |
Display | PC | Operating system | Peak Detection & System Operation |
Our 7'' Color VGA display (640 x 480 resolution) is easy to read. The display is a touch screen. | Embedded Pentium Computer with > 1 GByte of RAM; data storage on 1 Gbyte USB memory stick. | Windows XP® embedded stored on 2.5 GByte smart card (the PeakWorks program and all data are on the memory stick. | Controlled by our 32 bit PeakWorks software program with more than 20 years of development. |
ADC | Autoranging | Autozero | Programmable Voltage/current output |
Our 16 bit ADCs provide a resolution of 1 part in 64,000 and our signal algorithm minimizes noise by signal averaging in the msec range; the ADC can acquire positive or negative signals. | The 312 has 3 ranges with computer controlled autoranging- x1, x 10, x 100. | Beginning of each run. | 0-1VDC. |
Digital outputs | Battery life | Carrier gas | Programmable setpoints |
USB connection. | 3-4 hours; can operate off cigarette lighter adapter. | 50 liters at 1500 PSI; typical capillary column 15-20 cc/min; typical packed column 30-40 cc/min. | The low and high setpoints (depends on the options) can be customer programmed for each point and/or compound. |
Oven | |||
Size: Typical column diameter: 3.5'' Temperature range: Battery: 25-50 Degrees Celcius External power: 25-100 Degrees Celcius |