检出限MDL1 典型值:<0.5 pg(S)/sec
检出限 (SCD/FID串联) 典型值:<5.0 pg(S)/sec
典型灵敏度: g S/g C2 >2 × 107
线性3 >104
重复性<2% RSD 2 小时;<5% RSD 24小时
臭氧供气Ozone supply gas:干燥氧气Dry oxygen (Ultra Zero grade超零级)
氧化剂Oxidizer: 超零级空气,超零级氧气Ultra Zero grade air, Ultra Zero grade oxygen4
氢气:超零级
模拟输出: 0–1 V, 0–10 V (Standalone version only)
1 检出限MDL: Burner temperature 800 °C, 8 mL/min lower hydrogen, 38 mL/m upper hydrogen, 50 mL/min air, tert-butyl disulfide in SCD checkout sample (5190-7003) as the test compound, 30 m × 0.32 mm, 1 µm DB1 (123-1033), 50 °C for 3 minutes, 25 °C/min to 160 °C, hold 2 minutes, 1 µL splitless injection, fully stabilized burner.
2 选择性Selectivity: Defined as the sensitivity of S over the sensitivity of a selected hydrocarbon. Operating parameters same as MDL. Isooctane (sample solvent) and tert-butyl disulfide in the SCD checkout sample are used as the testing compounds. Selectivity performed on a fully stabilized burner.
3 线性Linearity: Operating parameters same as MDL except 12 mL/min lower hydrogen and 38 mL/m upper hydrogen; 80 °C for 1 minute, 25 °C/min to 160 °C, hold 2 minutes, COC inlet; Test compound: tert-butyl disulfide in isooctane. Linearity performed on a fully stabilized burner.
4 SCD specifications apply only when using air as the oxidizer gas.
硫化学发光检测器
重新构想和设计的 Agilent 8355 SCD 标志着硫化学发光检测技术 25 年来的第一次重大改进,使这项技术更加可靠且更易于使用。
Agilent 8355 SCD具有出色的灵敏度与特异性,且由于采用减少了 50% 组件的简化燃烧头设计而更易于维护。过去需要花费一小时的最常见服务程序如今仅需 10 分钟即可完成。
Agilent 8355 SCD 使用双等离子体燃烧器使含硫化合物在高温下燃烧生成一氧化硫 (SO)。光电倍增管可检测由 SO 和臭氧发生化学发光反应而产生的光。实现线性、等摩尔的硫化物响应,大部分样品基质都不会对其产生干扰。
特性:
用于气相色谱 (GC) 的硫特异性检测器
完全集成化的配置或独立的配置
皮克级检出限
没有烃的淬灭
对硫化物等摩尔线性响应
ASTM 方法兼容
串联 SCD 和 FID 操作
燃烧器组件减少约 40%;减少了潜在的泄漏点
更换内部陶瓷管仅需 10 分钟
安捷伦还提供 8255 氮化学发光检测器 (NCD)
除厂家/中国总经销商外,我们找不到 Agilent 8355 硫化学发光检测器 的一般经销商信息,有可能该产品在中国没有其它经销商。
如果您是,请告诉我们,我们的邮件地址是:sales@antpedia.net 请说明: 1.产品名称 2.公司介绍 3.联系方式 |
售后服务
我会维修/培训/做方法
如果您是一名工程师或者专业维修科学 仪器的服务商,都可参与登记,我们的平台 会为您的服务精确的定位并展示。