产品性能参数:
1、用于悬浮液和乳液表征;可以是水相,极性和非极性溶剂
2、超宽粒径测量范围,从 5nm 至 1000μm;单一理论
3、不用稀释,测量原浓样品:0.1~50%(体积百分数)
4、无需校正
5、频率范围:1 ~ 100 MHz
6、声频发生器与检测器之间的间隙可调
7、通过Access数据库进行数据管理
8、DT-110用于样品量小于5ml的粒度测定
Measurement Specifications
Physical Specifications
Sample Requirements | Models DT-1202/1210 particle size & zeta potential | Model DT-100/110 Particle size only | Model DT-600 Acoustic Longitudinal Rheometer | Model DT-300/310/330/350 Zeta Potential Probe | Model DT-700 Non-Aqueous Conductivity, Surfactant Content |
---|---|---|---|---|---|
Sample Volume Range | 0.1 -150 ml | 3 -70 ml | 3 -100 ml | 0.1-100 ml | 50 ml |
Volume Fraction Particulates (%) | 0.1-60 | 0.1-60 | no restriction | 0.1-60 | N/A |
Porosity (%) | 1 – 99 (Opt. 0014) | 1 – 99 (Opt. 0014) | N/A | 1 – 99 (Model 330/350) | N/A |
Conductivity | no restriction | no restriction | no restriction | no restriction | 1E-11 to 1E-04 |
pH | 0.5-13.5 | 0.5-13.5 | 0.5-13.5 | 0.5-13.5 | N/A |
Temperature [degree C] | below 50 | below 50 | below 50 | below 50 | below 100 |
Viscosity of media [cp] | to 20,000 | to 20,000 | to 20,000 | to 20,000 | to 20,000 |
Microviscosity of media [cp] | to 100 | to 100 | no restriction | to 100 | to 100 |
Viscosity of colloid | to 20,000 | to 20,000 | to 20,000 | to 20,000 | to 20,000 |
Particle size [microns] | 0.001 to 1000 | 0.001 to 1000 | no restriction | < 100, porous monolith | N/A |
Zeta potential | no restriction | no restriction | no restriction | no restriction | N/A |
Measured Parameters | |||||
Temperature [degree C] | 0 to 100, ±0.1 | 0 to 100, ±0.1 | 0 to 100, ±0.1 | 0 to 100, ±0.1 | N/A |
pH | 0.5-13.5, ±0.1 | 0.5-13.5, ±0.1 | 0.5-13.5, ±0.1 | 0.5-13.5, ±0.1 | N/A |
Frequency Range [MHz] | 1- 100 | 1- 100 | 1- 100 | 1- 100 | N/A |
Acoustic Attenuation [dB/cm MHz] | 0 to 20, ±0.01 | 0 to 20, ±0.01 | 0 to 20, ±0.01 | N/A | N/A |
Sound Speed [m/sec] | 500 to 3000,± 0.1 | 500 to 3000,± 0.1 | 500 to 3000,± 0.1 | N/A | N/A |
Electroacoustic signal | ±1% | N/A | N/A | ±1% | N/A |
Conductivity [S/m] (Options 004 & 0041) | 1E-11 to 10, ±1% | 1E-11 to 10, ±1% | N/A | 1E-11 to 10, ±1% | 1E-11 to 1E-4 |
Calculated Parameters | |||||
Particle size [microns] | 0.001 to 1000 | 0.001 to 1000 | N/A | N/A | N/A |
Unimodal parameters | Yes | Yes | N/A | N/A | N/A |
Bimodal parameters | Yes | Yes | N/A | N/A | N/A |
Zeta Potential [mV] | -200 to 200 ±(0.1) | N/A | N/A | -200 to 200 ±(0.1) | N/A |
Viscosity longitudinal [cP] | N/A | N/A | 0.5-20000,±3% | N/A | N/A |
Bulk viscosity for Newtonian liquids [cP] | N/A | N/A | 0.5-100,±3% | N/A | N/A |
Liquid compressibility [104/MPa-1] | N/A | N/A | 1-30, ±3% | N/A | N/A |
Newtonian liquid test at MHz range | N/A | N/A | N/A | N/A | N/A |
Porosity (%) | 1 – 99 (Opt. 0014) | 1 – 99 (Opt. 0014) | N/A | 1 – 99 (Model 330/350) | N/A |
Surfactant Content (%) | 1 – 99 (Opt. 0041) | 1 – 99 (Opt. 0041) | N/A | 1 – 99 (Opt. 0041) | 1 – 99 |
Measurement time [min] | |||||
Particle size | 1- 10 | 1- 10 | N/A | N/A | |
Zeta Potential in aqueous system | 0.5 | 0.5 | 0.5 | N/A | |
Zeta Potential in non-aqueous system< | 0.5-5 | N/A | 0.5-5 | 0.5-5 | N/A |
Rhelogy | N/A | N/A | 1-10 | N/A | N/A |
Weight [kg] | electronic unit: 20 sensor unit: 30 | electronic unit: 20 sensor unit: 30 | electronic unit: 20 sensor unit: 3 | electronic unit: 20 sensor unit: 7 | entire unit: 8 |
Power | 100-250 VAC 50-60 HZ 450 W | 100-250 VAC 50-60 HZ 450 W | 100-250 VAC 50-60 HZ 450 W | 100-250 VAC 50-60 HZ 450 W | 100-250 VAC 50-60 HZ 450 W |
Options Available with DT Instruments | |||||
Model | pH/Temp | Conductivity | Pump | Titration | Upgrade to DT- 1202 |
DT- 100 DT- 110 | Yes | Yes | Yes | Yes | Yes |
DT- 600 | Yes | Ye | Yes | Yes | Yes |
DT- 300 | Yes | Yes | Yes | Yes | No |
DT- 700 | No | Yes | Yes | No | No |
DT- 1202 DT- 1210 | Yes | Yes | Yes | Yes | N/A |
(1) Instrument can measure attenuation spectra well beyond 60 vol %, but theory for computing PSD from this data is limited to 60 vol %. The isoelectric pH is accurate over the entire range, however, some decrease in the absolute value of zeta potential may be experienced as one approaches the volume fraction limits.(3) In calculating the particle size the important viscosity is that experienced by the particle as it moves in response to the sound wave. This “microviscosity” in cases such as gels or other structured systems may be significantly less than the viscosity of the media measured with conventional rheometers which measure viscosity over a scale much larger than the particle.(2) For titration experiments it may be necessary to use external circulation pump to provide adequate mixing of reagents for quite viscous samples.(4) Particle size range for zeta potential measurement might depend on the density contrast.Computer Hardware/ Software SpecificationsCPU Intel G3320TE, 2.3 GHz LGA1150Mother Board Intel ChipsetHard Disk SSD >= 60 GBVideo Output HDMI StandardMemory DDRL3-1600 4GBUSB 2.0 Three PortsOperating System Microsoft Windows 7 |