General parameters 常规参数
Measuring principle: Fourier transform infrared, FTIR
Performance:Simultaneous analysis of up to 50
gas compounds
Response time, T90: Typically < 120 s, depending on
the gas flow and measurement
Time
Operating temperature:5 - 30°C, non-condensing
air conditioning recommended
Storage temperature: -20 - 60°C, non-condensing
Power supply:100-115 or 230 V / 50 -60 Hz
Power consumption: 300 W max, continuous 150
测量原理:FT-IR(傅里叶变换红外光谱分析)
性能:可同时测量多达50种气体组分
响应时间T90:< 120 s,取决于气体流速和测量时间
工作温度:5 - 30°C,无冷凝,建议使用空调以保持恒温。
贮存温度:-20 - 60°C,无冷凝
电源:100-115v or 230 V / 50 -60 Hz
功率:最大功率为300W,连续工作功率为150W
Spectrometer 分光仪
Resolution: 4 cm-1
Scan frequency:10 scans / s
Detector:Peltier cooled MCT
Source: SiC, 1550 K
Beamsplitter:ZnSe
Window material: ZnSe
Wave number range: 500 - 4 200 cm-1
分辨率:4 cm-1
扫描频率:10 次/s
检测器:Peltier 制冷 MCT
红外光源:SiC, 1550 K
分束器:ZnSe
窗口材质:ZnSe
波数范围:500 - 4200 cm-1
Sample cell 样品室
Structure:Multi-pass, fixed path length 5.0 m
Standard material:100 % rhodium coated aluminium
Mirrors: Fixed, protected gold coating
Volume: 0.4 liters
Connectors: Inlet Swagelok 6 mm
Outlet Swagelok 8 mm
Gaskets:Viton® O-rings
Temperature:180 °C, maximum
Window material: BaF2
结构:多册反射光程,固定路径长度为5.0m
标准材质:100%铑镀铝
反射镜:固定,黄金镀层保护
容积:0.4L
连接器:入口接头套管6mm,出口接头套管8mm;
垫片:O型氟化橡胶
温度:最高180℃
窗口材质:BaF2
Measuring parameters 测量参数
Zero point calibration: 24 hours, calibration with nitrogen (5.0 or higher N2 recommended)
Zero point drift:< 2 % of measuring range per zero point calibration interval
Sensitivity drift:None
Linearity deviation:< 2 % of measuring range
Temperature drifts:< 2 % of measuring range per 10K temperature change
Pressure influence:1 % change of measuring value for 1 % sample pressure change.Ambient pressure changes measured and compensated
零点校准:24H,用纯N2校准(推荐用5.0或更高的N2)
零点漂移:< 2 %(小于每个零点的校准区间量程< 2 %)
灵敏度漂移:无
线性偏差:< 2 %(测量量程的2 %);
温度漂移:每10K的温度变化内漂移量<量程2%
压力影响:样气压力变化1%,测量值变化1%,环境压力变化可被监测并得以补偿。
Electrical connectors 电连接器
Digital interface: 9-pole D-connector for RS-232 Analyzer is connected to an external computer via RS-232C cable. The external computer controls Gasmet.
Power connection:Standard plug CEE-22
数字接口:RS-232 D型9孔,分析仪通过 RS-232C 连接到外部计算机,由外部计算机控制傅立叶分析仪。
电源连接:CEE-22标准插头
Gas inlet and outlet conditions 气体进出口条件
Gas temperature:Non-condensing, the sample gas temperature should be the same as the sample cell temperature
Flow rate:120 - 600 liters per hour
Gas filtration:Filtration of particulates (2 μm) required
Sample gas pressure:Ambient
Sample pump:External, not included
气体温度:无凝结,样气温度必须和样品室的温度相同
流速:120-600L/h
气体过滤器:过滤精度2 μm
样气压力:大气压
样气泵:外置,不包含
Electronics 电气
A/D converter:Dynamic range 95 dB
Signal processor:32-bit floating point DSP 120 MFLOPS
Computer:External, not included
A/D转换器:动态范围:95 dB
信号处理器:32 bit浮点 DSP 120 MFLOPS
计算机:外置,不包含
Analysis software (for external PC)
Operating system:Windows 7 (32-bit)
Analysis software:Calcmet for Windows
分析软件(外部计算机)
操作系统:Windows 7 (32位)
分析软件:基于Windows的Calcmet
Options 选项
Pressure measurement: Inside sample cell
External PC:
Gasmet PC for control and analysis with optional alarm relay boards and analog inputs or outputs
Sample cell gaskets: Kalrez®
压力测量:在样品室内部
外置计算机:Gasmet工业电脑配备警报继电器板和模拟信号输入输出接口,用于系统控制和信号分析。
Enclosure 附件
Sample cell material: Aluminium
Dimensions (mm):482 * 196 * 450
Weight: 17 kg
CE label:According to EMI guideline 89/336/EC
样品室材质:铝
尺寸(mm):482 * 196 * 450
重量:17KG
CE标签:据 EMI 89/336/ec 号指令