技术参数
RH 范围 | 0 to 100 % |
RH 分辨率 | 1/264 % (RH probe) |
温度分辨率 | 1/264 °C (RH probe) |
RH 精确度 | RH probe sensor dependent (±1%, ±1.5%) |
露点 | Using RH probe for dew points above 80 °C is not recommended |
饱和器最max温度 | 90°C |
热保护 | 饱和器加热器的热切断。温度控制器中也设置温度限制。 |
流量 | 由MFC的范围决定,标准单位的可用范围:50, 100, 200,500 cc/min |
气体类型 | 惰性气体,通常为空气或者N2 |
进气口 | 1/8” (Swagelok® type bulkhead) |
进气口压力 | Maximum: 20 bar (300 psi) Minimum: 0.7 bar (10 psi) |
出气口
| 0.25” (6.35 mm) OD tubing –(the compression fitting can beeasily replaced to accommodate other sizes). |
传输管线尺寸 | Heated length about – typically 1 m (40”) Inner tube: 0.25” OD; 3/16” or 1/8” ID(6mm OD, 4 mm ID, optional) Thickness w/ insulation: about 19-20 mm (0.8”) Other lengths and sizes available as options |
传输管线温度: | Maximum (Continuous) 100 °C Minimum: Ambient |
尺寸
| (22.3 cm x 29.2 cm x 40.6 cm) (8.75” x 11.5” x 16”) |
仪器重量 | 13 kg (28 lb) |
电源要求
| 100-120 VAC, 300VA, 60 Hz nominal or: 220-240 VAC, 300VA, 50 Hz nominal |