Table-top Test Frame台式 | 台式 | 台式 | 立式 | 立式Floor Test Frame | ||
---|---|---|---|---|---|---|
Test Frame Type and Name | AGS-X 10 kN | AGS-X 20 kN | AGS-X 50 kN | AGS-X 100 kN | AGS-X 300 kN | |
Test Frame Load Capacity | Max. 10 kN (2,200 lbs) | Max. 20 kN (4,400 lbs) | Max. 50 kN (11,200 lbs) | Max. 100 kN (22,400 lbs) | Max. 300 kN (67,400 lbs) | |
Crosshead Speed Range | 0.001 to 1000 mm/min (0.0004 to 39 in/min) | 0.001 to 1600 mm/min (0.0004 to 62 in/min) | 0.001 to 800 mm/min (0.0004 to 31 in/min) | 0.001 to 500 mm/min (0.0004 to 19 in/min) | ||
Maximum Return Speed | 1500 mm/min (59 in/min) | 2200 mm/min (86 in/min) | 1100 mm/min (43 in/min) | 550 mm/min (21 in/min) | ||
Crosshead Speed Accuracy | Within ±0.1% of test speed | |||||
Crosshead Positional Accuracy | ±0.1% of indicated value or ±0.01mm, whichever is larger | |||||
Test Force Measurement Accuracy | High-Accuracy Load Cells | Within ± 0.5% of indicated test force (for 1/1 to 1/500 of load cell rated capacity) | Within ± 0.5% of indicated test force (for 1/1 to 1/250 of load cell rated capacity) | |||
Conforms to ISO 7500-1 Class 0.5, ASTM E4, JIS B7721 Class 0.5, EN 10002-2 Grade 0.5, BS1610 Class 0.5, and DIN51221 Class 0.5 | ||||||
Standard-Accuracy Load Cells | Within ± 1% of indicated test force (for 1/1 to 1/500 of load cell rated capacity) Conforms to ISO 7500-1 Class 1, ASTM E4, JIS B7721 Class 1, EN 10002-2 Grade 1, BS1610 Class 1, and DIN51221 Class 1 | |||||
Effective Width | 425 mm (16.5 in) | 600 mm (23.5 in) | ||||
Crosshead-Travel for Standard Height Models | Max. 1200 mm (47 in) | Max. 1250 mm (49 in) | Max. 1210 mm (48 in) | Max. 1255 mm (49 in) | Max. 1475 mm (58 in) | |
Crosshead-Travel for +250mm Extended Models | Max. 1450 mm (57 in) | Max. 1500 mm (59 in) | Max. 1460 mm (58 in) | Max. 1505 mm (59 in) | Max. 1725 mm (68 in) | |
Crosshead-Travel for +500mm Extended Models | Max. 1700 mm (67 in) | Max. 1750 mm (69 in) | Max. 1710 mm (68 in) | Max. 1755 mm (69 in) | Max. 1975 mm (78 in) | |
Data Capture Rate | 1000 Hz | |||||
Standard Input/Output Channels | –1 Dedicated load channel –1 Digital input/output channel*4 –1 Control input/output channel*4 | |||||
Dimensions | W653 x D520 x H1603 mm (W26 x D21 x H63in) | W718 x D641 x H1633 mm (W28 x D25 x H64 in) | W945 x D725 x H2164 mm (W37 x D29 x H85 in) | W945 x D725 x H2414 mm (W37 x D29 x H95 in | ||
Weight | Approx. 85 kg (187 lbs) | Approx. 235 kg (518 lbs) | Approx. 260 kg (573 lbs) | Approx. 525 kg (1157 lbs) | Approx. 675 kg (1488 lbs) | |
Operating Environment | Temperature: 5° C to 50° C (41° F to 122° F); Humidity: 20% to 80% (no condensation) Floor Vibrations: frequency 10 Hz max., amplitude 5 µm (0.0002 in) max. | |||||
Power Requirements | Single Phase 100/120/220/240 V 50/60 Hz | Single Phase 200-230 V 50/60 Hz | Single Phase 200-230 V 50/60 Hz | Three Phase 200-230 V 50/60 Hz *3 | Three Phase 200-230 V 50/60 Hz *3 |
Note 1: When the load cell capacity is smaller than the tester load capacity, the former is the maximum test force.
Note 2: Shimadzu recommends validation at an installation site that meets the requirements specified in these standards.
Note 3: Three Phase 380-440 V 50/60 Hz are available upon request for AGS-X 100kN and AGS-X 300kN Models.
Note 4: Optional expansion boxes can be used to increase the number of channels, see brochure for details.
PC Requirements for Trapezium X
Operating System | Windows 7 | |
Type | 32 or 64 bit | |
CPU | Intel Pentium 4, 2GHz or higher | |
Memory | 1 GB min. | |
HDD | 80 GB or greater | |
CD-ROM Drive | Required for software installation | |
USB Port | 1 port for main frame, other accessories may require additonal ports | |
Display | Size | 19 inch minimum |
Resolution | 1024 x 768 pixels minimum |
Installation Space
For Research Use Only. Not for use in diagnostic procedures.
规 格:
载荷容量:100KN、300KN
载荷精度:显示值的±0.5%(精度保证范围:0.2~100%)
试验速度:0.001~800mm/min(100kN)、0.001~400mm/min(300kN)
速度精度:±0.1%
返回速度:1100 mm/min(100kN)、550 mm/min(300kN)
位置检测分辨率:0.033μm
位置控制精度:±0.1%
数据采样间隔:1ms
除厂家/中国总经销商外,我们找不到 岛津AGS-X电子万能试验机(立式) 的一般经销商信息,有可能该产品在中国没有其它经销商。
如果您是,请告诉我们,我们的邮件地址是:sales@antpedia.net 请说明: 1.产品名称 2.公司介绍 3.联系方式 |
售后服务
我会维修/培训/做方法
如果您是一名工程师或者专业维修科学 仪器的服务商,都可参与登记,我们的平台 会为您的服务精确的定位并展示。