AbC™ Anti-Mouse Bead Kit

2019.4.22

实验概要

The  AbC™ Anti-Mouse Bead Kit provides a consistent, accurate, and  simple-to-use technique for the setting of flow cytometry compensation  when using fluorochrome-conjugated mouse antibodies (Picture 1). The kit  contains two types of specially modified polystyrene microspheres, the  AbC™ capture beads, that bind all isotypes of mouse immunoglobulin, and  the negative beads that have no antibody binding capacity.

After  incubation with a fluorochrome-conjugated mouse antibody, the two bead  components provide distinct positive and negative populations of beads  that you can use to set compensation. You can perform compensation with  the same fluorochromelabeled antibody used for cell staining. Because of  the consistent nature of bead scatter and high surface antibody binding  capacity, this allows you to more consistently and accurately set  compensation for any combination of fluorochrome-labeled mouse  antibodies.
 
 Picture  1. Compensation using the AbC™ Anti-Mouse Bead Kit. (A) Phycoerythrin  (PE)-conjugated mouse anti-human CD56 antibody (MHCD56044) labeling the  AbC™ capture beads for a positive signal and negative beads providing a  negative signal. (B) FITC-conjugated mouse anti-human CD3 (MHCD03014)  labeling the AbC™ capture beads for a positive signal and negative beads  providing a negative signal. (C) Dual parameter plot showing gated  humanlymphocytes labeled with Phycoerythrin-conjugated mouse anti-human  CD56 and FITC-conjugated mouse anti-human CD3 using compensation  settings obtained with the AbC™ Anti-Mouse Bead Kit.

 

实验步骤

1.  Using the AbC™ Anti-Mouse Bead Kit

1)         Completely resuspend the AbC™ capture beads (Component A) and negative  beads (Component B) by gently vortexing for 30 seconds before use.

2)         Label a sample tube for each fluorochrome-conjugated mouse antibody you  are using, and add 1 drop of AbC™ capture beads (Component A) to each  tube.

3)         Add a pre-titrated amount of each mouse antibody conjugate to the AbC™  capture bead suspension in the designated tube and mix well. Make sure  to deposit the antibody directly to the bead suspension.

4)        Incubate for 15 minutes at room temperature, protected from light.

5)        Add 3 mL of PBS or other buffer to sample tubes. Centrifuge for 5 minutes at 200 × g.

6)         Carefully remove the supernatant from tubes and resuspend the bead  pellet by adding 0.5 mL of PBS or other buffer to sample tubes.

7)        Add one drop of negative beads (Component B) to the tubes and mix well.

8)         Vortex tubes before analyzing using flow cytometry. You may briefly  sonicate to increase the percentage of singlet beads, if necessary.

9)         Perform manual or automatic compensation according to the preferred  procedure for the flow cytometer in use. Gate on the bead singlet  population based on FCS and SSC characteristics.

 

2.  Combining AbC™ and ArC™ Kits

The  ArC™ Amine Reactive Compensation Bead Kit is designed to facilitate  compensation when using any of the LIVE/DEAD® fixable dead cell stains,  all amine-reactive dyes. This kit provides two types of specially  modified polystyrene microspheres, the ArC™ reactive beads (Component A)  that bind any of the amine-reactive dyes, and the ArC™ negative beads  (Component B), that have no reactivity. Using the two kit components  with any aminereactive dye will provide distinct positive and negative  populations of beads. You can use the AbC™ Anti-Mouse Bead Kit and the  ArC™ Amine Reactive Compensation Bead Kit together to calculate  compensation in multicolor immunophenotyping experiments that  incorporate a LIVE/DEAD® fixable dye by following the protocol outlined  below:

1)         Gently vortex the ArC™ Amine Reactive Beads Kit and the AbC™ Anti-Mouse  Bead Kit components for 30 seconds to completely resuspend before use.

2)         Label a sample tube for the amine-reactive dye you are using, and add 1  drop of ArC™ reactive beads (Component A) to a labeled sample tube.  Allow ArC™ reactive beads to sit in the tube for 5 minutes to warm to  room temperature.

3)         Prepare fluorescent reactive dye according to kit instructions included  in the LIVE/DEAD® Fixable Dead Cell Stain Kit. For optimal performance  of ArC™ reactive beads, always use freshly prepared amine-reactive dye.  Do not use previously frozen dye solution.

4)         Add the proper amount of LIVE/DEAD® fixable dead cell stain to the bead  suspension and mix well. Make sure to deposit the amine-reactive dye  directly to the bead suspension.

5)         Label another sample tube for each fluorochrome-conjugated antibody you  are using, and add 1 drop of AbC™ capture beads (Component A in the  AbC™ Anti-Mouse Bead Kit) to each tube.

6)         Add a pre-titrated amount of antibody conjugate to the appropriate tube  and mix well. Make sure to deposit the antibody directly to the bead  suspension.

7)        Incubate for 30 minutes at room temperature, protected from light.

8)        Add 3 mL of PBS or other buffer to each sample tube. Centrifuge at 300 × g for 5 minutes to collect the beads.

9)         Carefully remove all supernatant from each tube. If using the red  fluorescent reactive dye (Cat. no. L23102), repeat step 8 for that tube.

10)    Resuspend the bead pellet by adding 0.5 mL of buffer to each sample tube.

11)     Add one drop of negative beads (Component B in the AbC™ Anti-Mouse Bead  Kit) to sample tube(s) containing the AbC™ capture beads.

12)     Add one drop of ArC™ negative beads (Component B in the ArC™ Amine  Reactive Compensation Bead Kit ) to sample tube(s) containing the ArC™  reactive beads.

13)    Vortex tubes before analyzing using flow cytometry.

14)     Perform manual or automatic compensation according to the preferred  procedure for the flow cytometer in use. Gate on the bead singlet  population based on FSC and SSC characteristics.