ASTM E2529-06e1
测试拉曼光谱仪分辨率的标准指南

Standard Guide for Testing the Resolution of a Raman Spectrometer


标准号
ASTM E2529-06e1
发布
2006年
发布单位
美国材料与试验协会
替代标准
ASTM E2529-06(2014)
当前最新
ASTM E2529-06(2022)
 
 
引用标准
ANSI Z136.1 ASTM E131 ASTM E1683 ASTM E1840
适用范围

4.1 Assessment of the spectrometer resolution and instrument line shape (ILS) function of a Raman spectrometer is important for intercomparability of spectra obtained among widely varying spectrometer systems, if spectra are to be transferred among systems, if various sampling accessories are to be used, or if the spectrometer can be operated at more than one laser excitation wavelength.

4.2 Low-pressure discharge lamps (pen lamps such as mercury, argon, or neon) provide a low-cost means to provide both resolution and wave number calibration for a variety of Raman systems over an extended wavelength range.

4.3 There are several disadvantages in the use of emission lines for this purpose, however.

4.3.1 First, it may be difficult to align the lamps properly with the sample position leading to distortion of the line, especially if the entrance slit of the spectrometer is underfilled or not symmetrically illuminated.

4.3.2 Second, many of the emission sources have highly dense spectra that may complicate both resolution and wave number calibration, especially on low-resolution systems.

4.3.3 Third, a significant contributor to line broadening of Raman spectral features may be the excitation laser line width itself, a component that is not assessed when evaluating the spectrometer resolution with pen lamps.

4.3.4 An alternative would use a Raman active compound in place of the emission source. This compound should be chemically inert, stable, and safe and ideally should provide Raman bands that are evenly distributed from 0 cm-1 (Raman shift) to the C-H stretching region 3000 cm -1 and above. These Raman bands should be of varying bandwidth.

4.4 To date, no such ideal sample has been identified; however carbon tetrachloride (see Practice E1683) and naphthalene (see Guide E1840) have been used previously for both resolution and Raman shift calibration.

4.5 The use of calcite to assess the resolution of a Raman system will be addressed in this guide. Calcite is a naturally occurring mineral that possesses many of the desired optical properties for a Raman resolution standard and is inexpensive, safe, and readily available.

4.6 The spectral bandwidth of dispersive Raman spectrometers is determined primarily by the focal length of the spectrometer, the dispersion of the grating, and the slit width. Field portable systems typically operate with fixed slits and gratings and thus operate with a fixed spectral bandwidth, while in many laboratory systems the slit widths and gratings are variable. The spectral bandwidth of Fourier-Transform (FT)-Raman systems is continuously variable by altering the optical path difference of the interferometer and furthermore is capable of obtaining much lower spectral bandwidth than most practical dispersive systems. Therefore, data obtained of a narrow Raman band on a FT-Raman system can be used to determine the resolution of a dispersive Raman ......


ASTM E2529-06e1相似标准


推荐

重庆研究院小型光谱仪样机研制成功

最后,通过选用光学探头与光栅光谱仪匹配设计,便携式光谱仪实际光谱分辨率可达10cm-1。  目前,重庆研究院已完成多种外形尺寸便携式和手持式光谱仪原理样机研制。这些样机对若干样品测试结果与标准库数据一致,从原理和技术上证实了小型光谱仪设计合理性和使用可靠性,下一步,团队研究重点将放在体积更小、性能更优微型光谱仪上,并开展小型光谱仪工程化和产业化应用。...

便携式光谱仪国内外技术PK

由福建计量院、厦门大学、厦门市普识纳米科技有限公司起草地方标准《便携式光谱快速检测仪》和《光谱仪》国家标准,将于2018年完成编订,该标准编订完成将为光谱仪生产、使用和检验提供技术依据,推动光谱技术发展,也规范市场应用,也为国产起到规范进步指导。  (四川省分析测试服务中心、四川省科学器材公司)...

药物成分分析

被测样品放置在标准、透明硼硅酸盐玻璃瓶中即可,而不需要进行其他多余样品制备。图1:对乙酰氨基酚(常称为扑热息痛镇痛药)光谱。使用模块化系统对样品进行分析,该系统包括QE系列光谱仪、500 mW输出785 nm激光器和光纤探头。 光谱仪波长范围为780-940 nm,为提高光学分辨率,配备50 μm狭缝。 同时加入高反射率反射镜,以提高光谱仪灵敏度。...

必达泰克将在Pittcon 2019发布新型手持光谱仪

NanoRam-1064完全符合所有管理法规,包括21 CFR Part 11和Part 1040.10,并在符合cGMP标准设施中发挥着不可或缺作用。 NanoRam-1064符合光谱法要求,包括美国药典,欧洲药典2.2.48,以及中国药典光谱学指南技术是一种公认、符合PIC/S和GMP指南关于原料100%身份保证方法。...


ASTM E2529-06e1 中可能用到的仪器设备





Copyright ©2007-2022 ANTPEDIA, All Rights Reserved
京ICP备07018254号 京公网安备1101085018 电信与信息服务业务经营许可证:京ICP证110310号