N91 地球科学仪器 标准查询与下载



共找到 95 条与 地球科学仪器 相关的标准,共 7

本标准规定了微机械系统加速度检波器的术语和定义、标记、要求、试验方法、检验规则、标志、包装、运输、贮存。 本标准适用于地震勘探用微机械系统加速度检波器。

Microelectro mechanical system acceleration geophone

ICS
25.040.40
CCS
N91
发布
2016-04-25
实施
2017-05-01

本标准规定了地震加速度检波器(以下简称“加速度检波器”)的术语和定义、组成、型式、分类与标记、要求、试验方法、检验规则、标志、包装、运输、贮存。 本标准适用于地震勘探用动圈式、涡流式、压电式加速度检波器。

Seismic acceleration geophone

ICS
75.180.10
CCS
N91
发布
2016-04-25
实施
2017-05-01

1.1 This practice covers the description, application, selection, installation, data collecting, and data reduction of the various types of extensometers used in the field of rock mechanics. 1.2 Limitations of each type of extensometer system are covered in Section 5. 1.3 The values stated in inch-pound units are to be regarded as the standard. The SI values given in parentheses are provided for information purposes only. 1.4 The text of this standard references notes and footnotes which provide explanatory material. These notes and footnotes (excluding those in tables and figures) shall not be considered as requirements of the standard. 1.5 This practice offers a set of instructions for performing one or more specific operations. This document cannot replace education or experience and should be used in conjunction with professional judgement. Not all aspects of this guide may be applicable in all circumstances. This ASTM standard i......

Standard Practice for Extensometers Used in Rock

ICS
93.020 (Earth works. Excavations. Foundation const
CCS
N91
发布
2012
实施

本规程适用于光学经纬仪的首次检定、后续检定和使用中检查。

Verification Regulation of Optical Theodolites

ICS
CCS
N91
发布
2011-11-14
实施
2012-05-14

本规程适用于经纬仪检定装置的首次检定、后续检定和使用中检查。

Verification Regulation of Theodolite Verification Devices

ICS
CCS
N91
发布
2011-11-14
实施
2012-05-14

本型式评价大纲适用于准确度等级为Ⅰ、Ⅱ、Ⅲ、Ⅳ级的电子经纬仪的型式评价。

Program of Pattern Evaluation for Electronic Theodolites

ICS
CCS
N91
发布
2011-11-14
实施
2012-02-14

Verification Regulations for Gyro Theodolite

ICS
CCS
N91
发布
2011-01-01
实施
2011-01-01

本标准规定了地震观测仪器的分类原则与方法、编码方法,给出了地震观测仪器分类代码。 本标准适用于地震观测仪器信息的管理和应用。

Classincation and code for earthquake observation instrument

ICS
17.040.30
CCS
N91
发布
2008-04-23
实施
2008-08-01

本标准规定了地震观测仪器(以下简称仪器)质量的检验规则,包括检验分类、检验项目、检验方法和判定规则。 本标准适用于各类地震观测仪器的质量检验。

Quality inspection provision for earthquake observation instrument

ICS
17.040.30
CCS
N91
发布
2008-04-23
实施
2008-08-01

本标准规定了地震观测仪器进网常用技术参数的定义、表述及测试方法。 本标准适用于地震观测仪器的设计、生产、使用、维护、引进和质量监督。

Technical requirements of instruments in network for earthquake monitoring.The description of common technical parmneter and test method

ICS
17.040.30
CCS
N91
发布
2007-03-14
实施
2007-06-01

本标准规定了地震观测仪器中地震仪进网的功能要求、技术指标、测试方法及环境适用性。 本标准适用于地震仪的设计、生产、使用、维护、引进和质量监督。

Technical requirements of instruments in network for earthquake monitoring.Seismograph

ICS
17.040.30
CCS
N91
发布
2007-03-14
实施
2007-06-01

本标准规定了地震观测仪器中陆地型系列重力仪的使用条件、功能要求、技术指标和检测方法。 本标准适用于重力仪的研制、生产、使用、维护、引进和质量监督。

Technical requirements of instruments in network for earthquake monitoring.Gravimeter

ICS
17.040.30
CCS
N91
发布
2007-03-14
实施
2007-06-01

1.1 This test method covers an integration of previous Test Method E 913 dealing with the calibration of pyranometers with axis vertical and previous Test Method E 941 on calibration of pyranometers with axis tilted. This amalgamation of the two methods essentially harmonizes the methodology with ISO 9846.1.2 This test method is applicable to all pyranometers regardless of the radiation receptor employed, and is applicable to pyranometers in horizontal as well as tilted positions.1.3 This test method is mandatory for the calibration of all secondary standard pyranometers as defined by the World Meteorological Organization (WMO) and ISO 9060, and for any pyranometer used as a reference pyranometer in the transfer of calibration using Test Method E 842.1.4 Two types of calibrations are covered: Type I calibrations employ a self-calibrating, absolute pyrheliometer, and Type II calibrations employ a secondary reference pyrheliometer as the reference standard (secondary reference pyrheliometers are defined by WMO and ISO 9060).1.5 Calibrations of reference pyranometers may be performed by a method that makes use of either an altazimuth or equatorial tracking mount in which the axis of the radiometer''s radiation receptor is aligned with the sun during the shading disk test.1.6 The determination of the dependence of the calibration factor (calibration function) on variable parameters is called characterization. The characterization of pyranometers is not specifically covered by this method.1.7 This test method is applicable only to calibration procedures using the sun as the light source.This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

Standard Test Method for Calibration of a Pyranometer Using a Pyrheliometer

ICS
07.060 (Geology. Meteorology. Hydrology)
CCS
N91
发布
2005
实施

The pyranometer is a radiometer designed to measure the sum of directly solar radiation and sky radiation in such proportions as solar altitude, atmospheric conditions and cloud cover may produce. When tilted to the equator, by an angle β, pyranometers measure only hemispherical radiation falling in the plane of the radiation receptor. This test method represents the only practical means for calibration of a reference pyranometer. While the sun-trackers, the shading disk, the number of instantaneous readings, and the electronic display equipment used will vary from laboratory to laboratory, the method provides for the minimum acceptable conditions, procedures and techniques required. While, in theory, the choice of tilt angle (β) is unlimited, in practice, satisfactory precision is achieved over a range of tilt angles close to the zenith angles used in the field. The at-tilt calibration as performed in the tilted position relates to a specific tilted position and in this position requires no tilt correction. However, a tilt correction may be required to relate the calibration to other orientations, including axis vertical. Note 18212;WMO Fist Class pyranometers, or better, generally exhibit tilt errors of less than 1 % to tilts of 50° from the horizontal. Traceability of calibrations to the World Radiometric Reference (WRR) is achieved through comparison to a reference absolute pyrheliometer that is itself traceable to the WRR through one of the following: One of the International Pyrheliometric Comparisons (IPC) held in Davos, Switzerland since 1980 (IPC IV). See Refs (3-7). Any like intercomparison held in the United States, Canada or Mexico and sanctioned by the World Meteorological Organization as a Regional Intercomparison of Absolute Cavity Pyrheliometers. Intercomparison with any absolute cavity pyrheliometer that has participated in either and IPC or a WMO-sanctioned intercomparison within the past five years and which was found to be within ±0.4 % of the mean of all absolute pyrheliometers participating therein. The calibration method employed in this test method assumes that the accuracy of the values obtained are independent of time of year, with the constraints imposed and by the test instrument's temperature compensation circuit (neglecting cosine errors).1.1 This test method covers an integration of previous Test Method E913 dealing with the calibration of pyranometers with axis vertical and previous Test Method E941 on calibration of pyranometers with axis tilted. This amalgamation of the two methods essentially harmonizes the methodology with ISO 9846. 1.2 This test method is applicable to all pyranometers regardless of the radiation receptor employed, and is applicable to pyranometers in horizontal as well as tilted positions. 1.3 This test method is mandatory for the calibration of all secondary standard pyranometers as defined by the World Meteorological Organization (WMO) and ISO 9060, and for any pyranometer used as a reference pyranometer in the transfer of calibration using Test Method E842. 1.4 Two types of calibrations are covered: Type I calibrations employ a self-calibrating, absolute pyrheliometer, and Type II calibrations employ a secondary reference pyrheliometer as the reference standard (secondary reference pyrheliometers are defined by WMO and ISO 9060). 1.5 Calibrations of reference pyranometers may be performed by a method that makes use of either an altazimuth or equatorial tracking mount in which the axis of the radiometer's radiation receptor is aligned with the sun during the ......

Standard Test Method for Calibration of a Pyranometer Using a Pyrheliometer

ICS
07.060
CCS
N91
发布
2005
实施

本部分规定了地震前兆观测仪器(以下简称仪器)通信与控制的约定与命令,适用于地震前兆观测仪器和地震前兆观测仪器通信控制软件的设计与生产。

Earthquake precursor observation instrument.Part 2:Communication and control

ICS
17.040.30
CCS
N91
发布
2003-11-07
实施
2004-05-01

本标准规定了激光法岩心粒度分析仪的原理与组成、要求、试验方法、检验规则、标志、包装、运输和贮存。 本标准适用于仪器的制造、检验和质量评价。

Core particle sizer

ICS
CCS
N91
发布
2003-03-18
实施
2003-08-01

This CID covers two types of illuminated, theodolites with accessories, cases, and tripods.

THEODOLITES, SURVEYING, DIRECTION: THIRD ORDER

ICS
CCS
N91
发布
2003
实施

CID covers two types of illuminated theodolites with accessories, cases, and tripods.

THEODOLITES, SURVEYING, DIRECTION: SECOND ORDER

ICS
CCS
N91
发布
2003
实施

この規格は,測量機器用の伸縮三脚(以下,三脚という。)について規定する。

Tripods for surveying instruments

ICS
17.180.30
CCS
N91
发布
2002-07-20
实施

この規格は,測量機器用三脚取付部(以下,三脚取付部という。)の寸法について規定する。

Tripod connections for optical surveying instrument

ICS
17.180.30
CCS
N91
发布
2002-07-20
实施



Copyright ©2007-2022 ANTPEDIA, All Rights Reserved
京ICP备07018254号 京公网安备1101085018 电信与信息服务业务经营许可证:京ICP证110310号