杭州赛析科技销售国产液相色谱仪,安捷伦二手液相,岛津二手液相,waters二手液相,安捷伦色谱柱,紫外可见分光光度计,联系电话15990028737,QQ1239219580,www.htcee.com
waters2695液相报错:Bubble found on compression
上一篇 /
下一篇 2018-04-25 14:23:29/ 个人分类:二手waters液相
waters2695液相报错:Bubble found on compression
Alarm Pump Detected Bubble
waters2695液相出现泵内有气泡,提示报警
判断:
The solvent management system detected a "bubble" during its compression cycle. The criterion for detection
of a bubble is the failure of the pre-compression stroke to achieve system pressure within 20% of the working stroke.
Possible causes are:
@A gas bubble momentarily interrupted solvent flow.
@The solvent delivery system is operating at near its maximum pressure vs. flow envelope
@High gas loading of solvents. Check the degas or sparge system for adequate solvent conditioning.
我公司提供waters2695液相维修,硬件维修,waters2695主板维修,waters2695泵控板维修,
鱼雷泵驱维修,泵头、压力传感器等维修服务
咨询电话15990028737,马先生
导入论坛收藏
分享给好友
推荐到圈子
管理
举报
TAG: onbubblefoundcompression