waters2695液相报错:Cannot reach minimum pressure

上一篇 / 下一篇  2018-04-26 10:01:41/ 个人分类:二手waters液相

waters2695液相报错:Cannot reach minimum pressure

                          Alarm Pump Ramp Decay Pressure

判断:失压、压力低于泵系统最小压力值

 

The solvent management system could not generate the minimum pressure required for the static leak test. May be caused by:

@Air in the system

@Leaky pin plug or cap

@Leak in fluidics

@Excessive leak in check valves

@Excessive plunger seal leak

排空的时候容易出现这个报错问题,可以忽略。如果正常运行的时候出现这个报错,检查是否漏液,检查单向阀是否正常等

我公司提供waters2695液相维修,硬件维修,waters2695主板维修,waters2695泵控板维修,

鱼雷泵驱维修,泵头、压力传感器等维修服务

咨询电话15990028737,马先生