1 Scope
This document specifies a calibration procedure of the energy step and the energy scale for electron energy loss spectroscopy in (scanning) transmission
electron microscopes to an uncertainty of ±3 % for the energy range 0 eV to 3 000 eV.
This document is intended for electron energy loss spectroscopy with transmitted electrons
through sufficiently electron transparent samples, such as a thin foil sample, and