Full Description This document describes a standard method for determining particle number size distributions in ambient air in the size range from 10 nm to 800 nm at total concentrations up to approximately 105 cm-3 with a time resolution of a few minutes. The standard method is based on a Mobility Particle Size Spectrometer (MPSS) used with a bipolar diffusion charger and a Condensation Particle Counter (CPC) as the detector. The document describes the performance characteristics and minimum requirements of the instruments and equipment to be used, and describes sampling, operation, data processing and QA/QC procedures, including calibration.
Cross References:
ISO 27891:2015
CEN/TS 16976:2016
ISO 15900
ISO 3696:1987
ISO/IEC GUIDE 98-3:2008 Ed 1
All current amendments available at time of purchase are included with the purchase of this document.