共找到 317 条与 声学仪器与测震仪 相关的标准,共 22 页
The document specifies a standard reference zero for the scale of hearing treshold level applicable to pure-tone air conduction audiometers, in order to promote agreement and uniformity in the expression of hearing treshold level measurements throughout the world. It states the information in a form suitable for direct application to the calibration of audiometers, that is, in terms of the response of two different standard models of earphone measured on a coupler complying with IEC 60303 and in terms of other supra-aural earphones of models specified in 4.3 measured on an aritficial ear complying with IEC 60318.#,,#
Acoustics - Reference zero for the calibration of audiometric equipment - Part 1: Reference equivalent treshold sound pressure levels for pure tones and supra-aural earphones (ISO 389-1:1998); German version EN ISO 389-1:2000
Acoustics - Statistical distribution of hearing thresholds as a function of age.
This part of IEC 61094 specifies mechanical dimensions and certain electroacoustic characteristics for condenser microphones used as laboratory standards for the realization of the unit of sound pressure and for sound pressure measurements of the highest attainable accuracy. The specifications are intended to ensure that primary calibration by the reciprocity method can be readily carried out. This part also establishes a system for classifying laboratory standard condenser micro-phones into a number of types according to their dimensions and properties in order to facilitate the specification of calibration methods, the conducting of inter-laboratory comparisons involving the calibration of the same microphones in different laboratories, and the interchangeability of microphones in a given calibration system.
Measurement microphones - Part 1: Specifications for laboratory standard microphones
Building acoustics - Estimation of acoustic performance of buildings from the performance of elements - Part 3 : airborne sound insulation against outdoor sound.
1.1この規格は,騒音計のランダム入射感度レべルを求めるための自由音場測定による校正方法及び拡散音場感度レべルを求めるための拡散音場測定による校正方法について規定する。1.2この規格では,拡散音場感度レべルとランダム入射感度レべルを同等のものとして扱う。使用できる設備に応じて,どちらの校正方法を用いてもよい。1.3この規格に従い校正された結果は,音場に置かれた騒音計の構成に依存する。参考この規格で求める感度レべルは.音場に置かれる機器の寸法や幾何学的形状によって大きく変わる可能性がある。例えば,マイクロホンの着脱が可能な騒音計などでは,実際の使用時と同じ構成にして音場に置く。1.4この規格でいう騒音計は,騒音計を含めたすべての音響測定システムを意味する。
Electroacoustics -- Random incidence and diffuse field calibration of sound level meters
1.1 These practices cover test specimen mountings to be used during sound absorption tests performed in accordance with Test Method C423. 1.2 The values stated in SI units are to be regarded as the standard. The values given in parentheses are provided for information only.
Standard Practices for Mounting Test Specimens During Sound Absorption Tests
Acoustics. Guidelines for noise control by silencers.
The document specifies the following data applicable to the calibration of bone vibrators for pure-tone bone conduction audiometry: a) Reference equivalent threshold force levels, corresponding to the threshold of hearing of young otologically normal persons by borne-conduction audiometry. b) Essential characteristics of the bone vibrator and of its method of coupling to a person under test and to the mechanical coupler. c) Essential characteristics and datum level of the masking noise applied to the ear not under test.#,,#
Acoustics - Reference zero for the calibration of audiometric equipment - Part 3: Reference equivalent threshold force levels for pure tones and bone vibrators (ISO 389-3:1994, including Technical Corrigendum 1:1995); German version EN ISO 389-3:1998
Acoustics - Reference zero for the calibration of audiometric equipment - Part 4: Reference levels for narrow-band masking noise (ISO 389-4:1994); German version EN ISO 389-4:1998
This part of ISO 389 specifies a standard reference zero for the scale of hearing threshold level applicable to pure-tone air conduction audiometers, in order to promote agreement and uniformity in the expression of hearing threshold level measurements throughout the world. It states the information in a form suitable for direct application to the calibration of audiometers, that is, in terms of the response of two different standard models of earphone measured on a coupler complying with EC 60303 and in terms of other supra-aural earphones of models specified in 4.3 measured on an artificial ear complying with IEC 60318. It is based on an assessment of the information available from the various standardizing laboratories responsible for audiometric standards and from scientific publications. Some notes on the derivation and application of the recommended reference levels are given in annex A.
Acoustics - Reference zero for the calibration of audiometric equipment - Part 1: Reference equivalent threshold sound pressure levels for pure tones and supra-aural earphones
Vibration measuring instrumentation - Vocabulary
1.1 This practice provides the user with procedures used in manually calibrating the fathometer or electronic depth sounder. This narrative describes calibration terminology, describes acceptable environmental conditions for calibration, and describes the calibration procedures. 1.2 The references cited contain useful information in the construction and the correct operation of the calibration equipment. 1.3 Any references cited in this narrative to specific products or brand names are made for information only, and is intended to be descriptive, but not restrictive, of products that will perform satisfactorily. 1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.
Standard Practice for Calibrating a Fathometer Using a Bar Check Method
1.1 This practice covers requirements for the secondary calibration of acoustic emission (AE) sensors. The secondary calibration yields the frequency response of a sensor to waves of the type normally encountered in acoustic emission work. The source producing the signal used for the calibration is mounted on the same surface of the test block as the sensor under testing (SUT). Rayleigh waves are dominant under these conditions; the calibration results represent primarily the sensor''s sensitivity to Rayleigh waves. The sensitivity of the sensor is determined for excitation within the range of 100 kHz to 1 MHz. Sensitivity values are usually determined at frequencies approximately 10 kHz apart. The units of the calibration are volts per unit of mechanical input (displacement, velocity, or acceleration). 1.2 The values stated in SI units are to be regarded as the standard. The values given in parentheses are for information only. 1.3 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.
Standard Practice for Secondary Calibration of Acoustic Emission Sensors
1.1 This practice is recommended for use in testing and measuring operating characteristics of acoustic emission electronic components or units. (See Appendix X1 for a description of components and units.) It is not intended that this practice be used for routine checks of acoustic emission instrumentation, but rather for periodic calibration or in the event of a malfunction. The sensor is not addressed in this document other than suggesting methods for standardizing system gains (equalizing them channel to channel) when sensors are present. 1.2 Where the manufacturer provides testing and measuring details in an operating and maintenance manual, the manufacturer''s methods should be used in conjunction with the methods described in this practice. 1.3 Difficult or questionable instrumentation measurements should be referred to electronics engineering personnel. 1.4 The methods set forth in this practice are not intended to be either exclusive or exhaustive. 1.5 The methods (techniques) used for testing and measuring the components or units of acoustic emission instrumentation, and the results of such testing and measuring should be documented. Documentation should consist of photographs, charts or graphs, calculations, and tabulations where applicable. 1.6 AE systems that use mini or micro computers to control the collection, storage, display, and analysis of data are in common use. Features of the computer-based systems include a wide selection of measurement parameters relating to the AE event. This selection, however, is usually made after the data have been acquired. This implies that the AE signals are individually recorded for later analysis, or that all the available parameters are measured on every AE signal that exceeds the selected threshold. The latter is usually the case. The manufacturer provides a specification for each system that specifies the operating range and conditions for the system. All calibration and acceptance testing of computer-based AE systems must use the manufacturer''s specification as a guide. This practice does not cover testing of the computer or computer peripherals. 1.7 This standard does not purport to address all of the safety problems, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.
Standard Practice for Characterizing Acoustic Emission Instrumentation
The purpose of this practice is to enable the transfer of calibration from sensors that have been calibrated by primary calibration to other sensors.1.1 This practice covers requirements for the secondary calibration of acoustic emission (AE) sensors. The secondary calibration yields the frequency response of a sensor to waves of the type normally encountered in acoustic emission work. The source producing the signal used for the calibration is mounted on the same surface of the test block as the sensor under testing (SUT). Rayleigh waves are dominant under these conditions; the calibration results represent primarily the sensor''s sensitivity to Rayleigh waves. The sensitivity of the sensor is determined for excitation within the range of 100 kHz to 1 MHz. Sensitivity values are usually determined at frequencies approximately 10 kHz apart. The units of the calibration are volts per unit of mechanical input (displacement, velocity, or acceleration). 1.2 The values stated in SI units are to be regarded as the standard. The values given in parentheses are for information only. 1.3 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.
Standard Practice for Secondary Calibration of Acoustic Emission Sensors
Acoustics. Cartography of outside environment noise. Drawing up of maps and graphical representation.
Acoustics. Laboratory tests on noise emission from appliances and equipment used in water supply installations. Part 3 : mounting and operating conditions for in-line valves and appliances.
Gives recommendations for sound systems intended for communicating speech, music and other signals primarily for monaural systems. Simultaneous interpretation systems, some aspects for recording studios, broadcasting studios, and home entertainment systems are not covered by this standard.
Code of practice for the design, planning, installation, testing and maintenance of sound systems
Audiometers. Part 2 : equipment for speech audiometry.
Audiometers - Part 2: Equipment for speech audiometry (IEC 60645-2:1993); German version EN 60645-2:1997
Copyright ©2007-2022 ANTPEDIA, All Rights Reserved
京ICP备07018254号 京公网安备1101085018 电信与信息服务业务经营许可证:京ICP证110310号