用于体内药物/基因/蛋白质传递及基因治疗电极
Tweezertrodes电极是一种可复用的小钳状电极,主要用于体内的药物或基因的传递。电极是一个标准的11.5厘米的小钳子,其尖部为嵌入的不锈钢圆盘状电极,两盘的间隙可调至2厘米,并有一个阳极指示器。
Tweezertrodes电极有两种规格,45-0118为直径:7mm的电极盘,45-0119为直径:10mm的电极盘
当把感兴趣的分子注射到体内后,用电极盘夹住相关组织,然后给出电脉冲,使孔道开始形成,将分子整合到直接与电极盘接触的组织细胞内。
Tweezertrodes are reusable, tweezer-style in vivo electrodes for drug or gene delivery in animals. Tweezertrodes consist of a standard 11.5 cm tweezer that has been modified with stainless steel circular or disk electrodes at the tip. The gap between the electrode disks may be adjusted from under 1 mm to over 2 cm. Tweezertrodes come in two models. Model 520 has an electrode disk diameter of 7 mm and Model 522 has a diameter of 10 mm. Tweezertrodes may be cleaned with a mild detergent and sterilized with ethanol or ethylene oxide. These electrodes are connected to a pulse generator with the Model 524 Connection Cable, and are compatible with most BTX Electroporation System
Applications

Tweezertrodes may be used for many in vivo applications, including gene and drug delivery. Following localized or systemic injection of the molecule of interest, the Tweezertrode electrode disks are used to grasp the tissue of interest. An electroporation pulse is then given, initiating pore formation and incorporation of the molecule into the cells of the tissue in direct contact with the electrode disks. Suzuki et al. used Tweezertrodes to deliver a Green Fluorescent Protein (GFP) gene into rat liver. The authors demonstrated successful gene transfer and expression in situ.1,2 Saito and Nakatsuji used a similar electrode to introduce plasmid DNA during ex-utero and in-utero electroporation.7 Sato et al. used the Tweezertrode with the T820 to introduce EGFP expressing plasmid in mouse testis.8 Tweezertrodes complement BTX Genetrode Models 508, 510, Caliper Electrode Models 384 and 384L, 2-Needle Array Models 530 and 532 as the newest addition to our in vivo electrode line. The Tweezertrodes can be used to reproduce applications requiring the caliper electrodes, including gene therapy, transdermal drug delivery and electroporation therapy.3,4,5,6

Technical Specifications


Standard Capabilities
Voltage Range: 0V - 200V (Do not use AC current)
Pulse Length Range: 1 µsec - 200 msec
Physical Characteristics
Model 520
Tweezertrode Length: 11.5 cm
Electrode Diameter: 7 mm
Electrode Material: Stainless steel
Model 522
Tweezertrode Length: 11.5 cm
Electrode Diameter: 10 mm
Electrode Material: Stainless steel
Generator Compatibility ECM T820, 630, 830, 2001

Ordering Information

System Model Available Configurations
520 Tweezertrode 520
Tweezer Electrode with stainless steel plated 7 mm diameter electrode disk for in vivo applications

523 Tweezertrode 523
Tweezer Electrode with stainless steel plated 10mm diameter electrode disk for in vivo applications

Accessory Model Description
524 Tweezertrode Connection Cable

References
Suzuki et al., Direct Gene Transfer into Rat Liver Cells by in vivo Electroporation, FEBS Letters, 425: 436-440 (1998)
BTX Electroporation Protocol PR0363 (1998)
Zhang et al., In vivo transdermal delivery of large molecules by pressure-mediated electroincorporation and electroporation: a novel method for drug and gene delivery, Bioelectrochemistry and Bioenergetics, 42: 283-292 (1997)
Dev, Giordano and Brown, In vivo delivery of gene to rabbit carotid artery by electroporation, Third U.S. / Japan Symposium on Drug Delivery Systems (Abstr.), 38 (1995)
Dev, Killing cancer cells with a combination of pulsed electric fields and chemotherapeutic agents, Cancer Watch, 3: 12-14 (1994)
BTX Electroporation Therapy Protocols ECT001-ECT005 (1995)
Saito and Nakatsuji, Developmental Biology, 240: 237-246 (2001)
Sato et al., Molecular Reproduction and Development, 61: 49-56 (2002)

活体转基因--钳状电极信息由美国自然基因有限公司为您提供,如您想了解更多关于活体转基因--钳状电极报价、型号、参数等信息,欢迎来电或留言咨询。

注:该产品未在中华人民共和国食品药品监督管理部门申请医疗器械注册和备案,不可用于临床诊断或治疗等相关用途

活体转基因--钳状电极 - 产品推荐
移动版: 资讯 直播 仪器谱

Copyright ©2007-2023 ANTPEDIA, All Rights Reserved

京ICP备07018254号 京公网安备1101085018 电信与信息服务业务经营许可证:京ICP证110310号