工商注册信息已核实!参考报价: | 面议 | 型号: | Standard Types |
品牌: | incoatec | 产地: | 德国 |
关注度: | 15 | 信息完整度: | |
样本: | 典型用户: | 暂无 | |
产地类别 | 进口 | 供应商性质 | 总代理 |
组件类别 | 光学元件 |
组件类别 | 光学元件 |
400-6699-117转1000
LiF(200), LiF(220), LiF(420)
LiF crystal types exist in a variety of lattice planes (200, 220, 420, etc.). In the sequence (200) --> (220) --
> (420), resolution increases and reflectivity decreases (Fig. 17).
Fig. 17: Intensities of the crystals LiF(220) and LiF(420) in relation to LiF(200). (Intensity LiF(200) =
1)
LiF(200):
A universally usable crystal for the element range atomic number 19 (K) onwards; high reflectivity, high
sensitivity.
LiF(220):
Lower reflectivity than LiF(200) but higher resolution; can be used for the element range atomic number
23 (V) onwards; particularly suitable for better peak separation where peaks overlap.
Examples of the application of the LiF(220) for reducing peak overlaps:
Cr Ka1,2 - V Kb1
Mn Ka1,2 - Cr Kb1
U La1 - Rb Ka1,2
LiF(420):
One of the special crystals; can be used for the element range atomic number 28 (Ni or Co Kb1) onwards;
best resolution but low reflectivity.
Figure 17 shows a reflectivity of only 10% of that of LiF(200) for LiF(420) in the energy range around 10
keV.
PET:
A universal crystal for the elements Al to Ti (K-peaks), Kr to Xe (L-peaks) and Hf to Bi (M-peaks).
ATTENTION
The PET is the crystal with the greatest heat-expansion coefficients, i.e. temperature fluctuations are
most noticeable here.
Fundamental Principles Introduction to X-Ray Fluorescence (XRF)
标准型XRF分析晶体信息由 北京众星联恒科技有限公司为您提供,如您想了解更多关于标准型XRF分析晶体报价、型号、参数等信息,欢迎来电或留言咨询。
注:该产品未在中华人民共和国食品药品监督管理部门申请医疗器械注册和备案,不可用于临床诊断或治疗等相关用途