Throughput
Up to 450 photometric and electrolyte tests/hour.
Reagent System
64 refrigerated reagent positions for 10 mL, 25 mL and 50 mL reagent bottles
Cooled to 10°C +/- 4°C at normal laboratory conditions
All reagent positions can be assigned as R1, R2 and R3.
Sample System
80 barcode readable positions
All positions fit 13×75 or 16x 100mm primary and secondary tubes and pediatric cups
All positions can be used for calibrators, controls, normal, pediatric and STAT samples
Pipetting System
1000 µL reagent syringe:
R1 volume 110-400 µL
R2 volume 0-180 µL
R3 volume 0-180 µL
Programmable in 1 µL steps
100 µL sample syringe:
Sample volume 1-30 µL
Programmable in 0.1 µL steps
Pre programmed ratio selectable from ratios 1:2 up to 1:100 with 3 diluents
Pre-heated probe with level detection, collision protection and integrated mixer.
Cuvette Rotor
Cost effective, semi-disposable cuvette rotor with 48 cuvettes
Up to 10.000 tests per rotor
Typical reagent consumption 220 µL per test
Measurement temperature 37°C, controlled by Peltier elements.
Light Source
Quartz-iodine lamp 12V-20W
Wavelength Range
Optical unit with 8 position filter wheel
340, 405, 505, 546, 578, 620, 660, 700 nm standard wavelengths
Other wavelengths available on request.
Photometric Range
-0.1 to 3.0 Absorbance
Resolution 0.001 Abs.
Analytical Modes (Single, Dual and Triple reagent system)
Kinetic measurement with linearity check
Mono- and bichromatic end point measurement with or without bichromatic reagent blank and/or sample blank correction
Two point measurement; with or without slope blank
Graphic plot of all measuring points
Predilution and automatic reflex dilution as needed
Range of calibration models
Non-linear calibration curves
Prozone check for immunology tests
Cut-off declaration
Calculated tests.
Quality Control
Up to 15 controls can be defined, 3 per test
Westgard or Basic QC rules
Levey-Jennings plots
Quality control statistics.
Standards and Regulations
CE -IVD
CB
UL.
Dimensions
117 x 115 x 80 cm (W x H x D)
46.1 x 45.3 x 31.5 in(W x H x D)
Weight: 190 kg (419 lbs).
Interface
Host-Query interface available
Host: RS 232 or Ethernet (TCP/IP) through LIS-2A protocol.
Environmental Conditions
Temperature: 15 – 32 °C (59 – 90 °F)
Humidity: 15 – 88% RH
Altitude: up to 2000 m
Plumbing: No dedicated water or drain required
Electrical:
Voltage: 100 – 240 Vac
Frequency: 50 – 60 Hz
Power (max): 380 VA .
Integrated PC
Touch screen navigation
Operating system: Windows™ XP
Options
Barcode Reader
Hand held CCD barcode reader used for reagent identification
Eco Kit
Enables separate collection of concentrated measurement mixture and dilute (wash) waste
Dry ISE Module
Patented Solid State Electrode Technology
Indirect measurement
Dilution 1:14
Measures Sodium, Potassium, Chloride and CO2.
Positive Sample Identification
Positive Sample Identification (PSID) reads all popular barcodes.
Printer
Printer supported by MS Windows™.
ProActive Maintenance Kit
Complete parts kit for annual preventive maintenance.