Procedure III

Used by Denise Hardy, ARUP Laboratories, Salt Lake City, UT.

Pre-fix < 2 mm specimens in fixative, room temperature, 1 hour. Microwave in reagent alcohol, 67°C, 5 minutes. Microwave in 100% isopropyl alcohol, 74°C, 5 minutes. Microwave in paraffin, 80°C, 5 minutes (use preheated to 75°C).

 