Safety requirements for electrical equipment for measurement, control and laboratory use - Part 2-041: Particular requirements for autoclaves using steam for the treatment of medical materials, and for laboratory processes
AUTOCLAVES, LABORATORY EQUIPMENT, SAFETY REQUIREMENTS, LABORATORY WARE,ELECTRICAL MEASURING INSTRUMENTS, ELECTRIC CONTROL EQUIPMENT
被代替标准
2004-01-01
适用范围
This National Standard of Canada is equivalent to International Standard CEI/IEC 1010-2-041:1995. 1 Scope and object This clause of part 1 is applicable except as follows: 1.1 Scope Replacement: This standard applies to AUTOCLAVES, including those with an automatic loading
and unloading system, which incorporate a PRESSURE VESSEL using steam within
the absolute pressure range from 0 to 500 kPa, and intended for the treatment
of medical materials and for laboratory processes, for example for
sterilization. NOTEs 1 National and other regulations or codes apply for the safety of
automatic loading and unloading systems. 2 All pressures are specified in absolute terms. Atmospheric pressure
(1 bar) = 100 kPa. Where an AUTOCLAVE incorporates a steam generator having its own PRESSURe
VESSEL within the same enclosure, the applicable safety requirements for the
AUTOCLAVE PRESSURE VESSEL apply also to the steam generator (see 14.105). 1.1.2 Equipment excluded from the scope Addition: Add the following new dash: - pressurized environmental cabinets. NOTES 1 Since all AUTOCLAVES covered by this part 2 use steam, other types using dry
heat, toxic gas or radiation are not included. 2 This standard does not deal with special requirements for protection
against chemical and high-risk micro-biological hazards associated with the
AUTOCLAVE LOAD, nor with requirements for the design of the PRESSURE VESSEl
itself.