IEC 60731-1997
医用电气设备 放射治疗中使用的带电离室的剂量仪

Medical electrical equipment - Dosimeters with ionization chambers as used in radiotherapy


 

 

非常抱歉,我们暂时无法提供预览,您可以试试: 免费下载 IEC 60731-1997 前三页,或者稍后再访问。

点击下载后,生成下载文件时间比较长,请耐心等待......

 



标准号
IEC 60731-1997
发布日期
1997年07月
实施日期
废止日期
中国标准分类号
C43
国际标准分类号
17.240
发布单位
IX-IEC
代替标准
IEC 60731-2011
适用范围
1 This International Standard specifies the performance requirements of radiotherapy dosimeters, as defined in 3.1, intended for the measurement of ABSORBED DOSE TO WATER Or AIR KERMA (and their RATES) in photon or electron radiation fields as used in radiotherapy. NOTE - Throughout this standard: - if no material is specified, the term "ABSORBED DOSE" Or "DOSE" means "ADSORBED DOSE TO WATER (in water)' and the term "KERMA" means "AIR KERMA (in air)"; - when the quantity "AIR KERMA (in air)" in units "Gy" is used, the quantity "EXPOSURE" in units "C/kg" is also allowable. 2 The dose-monitoring systems incorporated in radiotherapy treatment machines are not covered by this standard, neither are the re-entrant ion chambers used for brachytherapy source calibration. 3 This standard is applicable to the following types of DOSIMETER: a) FIELD-CLASS DOSIMETERS normally used for 1) the measurement of KERMA or DOSE in a radiation beam, either in air or in a PHANTOM; 2) in vivo skin surface or intracavitary measurements of DOSE on PATIENTS. b) REFERENCE-CLASS DOSIMETERS normally used for the calibration of FIELD-CLASS DOSIMETERS. NOTE - REFERENCE-CLASS DOSIMETERS may be USed as FIELD-CLASS DOSIMETERS.