ASTM E2330-2004
用感应耦合等离子体质谱法(ICP-MS)测定玻璃样品中痕量元素的标准试验方法

Standard Test Method for Determination of Trace Elements in Glass Samples Using Inductively Coupled Plasma Mass Spectrometry (ICP-MS)


ASTM E2330-2004 中,可能用到以下仪器设备

 

Neoma  多接收器 ICP-MS

Neoma 多接收器 ICP-MS

赛默飞色谱与质谱

 

ASTM E2330-2004



标准号
ASTM E2330-2004
发布日期
2004年
实施日期
废止日期
中国标准分类号
Q33
国际标准分类号
71.040.40 (Chemical analysis); 81.040.10 (Raw mate
发布单位
美国材料与试验协会
适用范围

This technique is destructive, in that the glass fragments may need to be crushed, and must be digested in acid.

Although the concentration ranges of the calibration curves shown in Appendix X1 are applicable to soda lime and borosilicate glass, this method is useful for the accurate measurement of trace elements from a wide variety of glass samples.

The determination the elemental concentrations in glass allows for additional data that can be compared between fragments. A standardized, validated method can aid in the interchange of data between laboratories.

It should be recognized that the method measures the bulk concentration of the target elements. Any extraneous material present on the glass that is not removed before digestion may result in a measurably different concentration of the elements.

The precision and accuracy of the method should be established in each laboratory that employs the method. Confidence intervals or a similar statistical quality statement should be quoted along with any reported concentration value.

1.1 This test method covers a procedure for elemental analysis of magnesium (Mg), aluminum (Al, titanium (Ti), manganese (Mn), gallium (Ga), rubidium (Rb), strontium (Sr), zirconium (Zr), antimony (Sb), barium (Ba), lanthanum (La), cerium (Ce), samarium (Sm), hafnium (Hf) and lead (Pb) in the bulk of glass samples (irregularly shaped and as small as 200 micrograms) for the comparison of fragments of a known source to the recovered fragments from a questioned source. These elements are present in soda lime and borosilicate glass in ngg-1 to % levels. Alternative methods for the determination of elemental analysis of glass are listed in the references. This procedure may also be applicable to other elements, other types of glass, or other concentration ranges with appropriate modifications of the digestion procedure (if needed for full recovery of the additional elements) and the mass spectrometer conditions. The addition of calcium and potassium, for example, could be added to the list of analytes in a modified analysis scheme.

This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.