ASTM E925-2002
光谱狭缝宽度不超过2nm的紫外线-可见光分光光度计的校准监测的标准实施规程

Standard Practice for Monitoring the Calibration of Ultraviolet-Visible Spectrophotometers whose Spectral Slit Width does not Exceed 2 nm


ASTM E925-2002 中,可能用到以下仪器设备

 

AXIMA Performance – MALDI TOF/TOF 质谱仪

AXIMA Performance – MALDI TOF/TOF 质谱仪

岛津企业管理(中国)有限公司/岛津(香港)有限公司

 

Buchi步琦Glass Oven B-585 Kugelrohr (球管玻璃炉)

Buchi步琦Glass Oven B-585 Kugelrohr (球管玻璃炉)

瑞士步琦公司BUCHI Labortechnik AG

 

ASTM E925-2002



标准号
ASTM E925-2002
发布日期
2002年
实施日期
废止日期
中国标准分类号
N53
国际标准分类号
17.180.30 (Optical measuring instruments); 71.040.
发布单位
US-ASTM
适用范围

This practice permits an analyst to compare the performance of an instrument to the manufacturerrsquo;supplied performance specifications and to verify its suitability for continued routine use. It also provides generation of calibration monitoring data on a periodic basis, forming a base from which any changes in the performance of the instrument will be evident.

1.1 This practice covers the parameters of spectrophotometric performance that are critical for testing the adequacy of instrumentation for most routine tests and methods within the wavelength range of 200 to 700 nm and the absorbance range of 0 to 2. The recommended tests provide a measurement of the important parameters controlling results in spectrophotometric methods, but it is specifically not to be inferred that all factors in instrument performance are measured.

1.2 This practice may be used as a significant test of the performance of instruments for which the spectral slit width does not exceed 2 nm and for which the manufacturer''s specifications for wavelength and absorbance accuracy do not exceed the performance tolerances employed here. This practice employs an illustrative tolerance of 177; 1 % relative for the error of the absorbance scale over the range of 0.2 to 2.0, and of 1.0 nm for the error of the wavelength scale. A suggested maximum stray radiant power ratio of 4 x 10-4 yields <1 % absorbance bias at an absorbance of 2. These tolerances are chosen to be compatible with many chemical applications while comfortably exceeding the uncertainty of the certified values for the reference materials and typical manufacturer''s specifications for error in the wavelength and absorbance scales of the instrument under test. The user is encouraged to develop and use tolerance values more appropriate to the requirements of the end use application. This procedure is designed to verify quantitative performance on an ongoing basis and to compare one instrument''s performance with that of other similar units. Refer to Practice E 275 to extensively evaluate the performance of an instrument.

1.3 This practice should be performed on a periodic basis, the frequency of which depends on the physical environment within which the instrumentation is used. Thus, units handled roughly or used under adverse conditions (exposed to dust, chemical vapors, vibrations, or combinations thereof) should be tested more frequently than those not exposed to such conditions. This practice should also be performed after any significant repairs are made on a unit, such as those involving the optics, detector, or radiant energy source.

1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.