71.040.50 (Physicochemical methods of analysis) 标准查询与下载



共找到 357 条与 相关的标准,共 24

This method is applicable to the simultaneous determination of dissolved alkali and alkaline earth cations and ammonium in water and wastewaters. Alkali and alkaline earth cations are traditionally determined by using spectroscopic techniques, such as AAS or ICP; whereas ammonium can be measured by using a variety of wet chemical methods, including colorimetry, ammonia-selective electrode, and titrimetry. However, ion chromatography provides a relatively straightforward method for the simultaneous determination of cations, such as lithium, sodium, potassium, calcium, magnesium, and ammonium, in fewer than 20–30 min.1.1 This test method is valid for the simultaneous determination of the inorganic alkali and alkaline earth cations, lithium, sodium, potassium, magnesium, and calcium, as well as the ammonium cation in reagent water, drinking water, and wastewaters by suppressed and nonsuppressed ion chromatography.1.2 The anticipated range of the method is 0.05-200 mg/L. The specific concentration ranges tested for this method for each cation were as follows (measured in mg/L):Lithium0.4-10.0Sodium4.0-40.0Ammonium0.4-10.0Potassium1.2-20.0Magnesium2.4-20.0Calcium4.0-40.01.2.1 The upper limits may be extended by appropriate dilution or by the use of a smaller injection volume. In some cases, using a larger injection loop may extend the lower limits.1.3 It is the user's responsibility to ensure the validity of these test methods for waters of untested matrices.1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use. For hazards statements specific to this test method, see 8.3.

Standard Test Method for Determination of Dissolved Alkali and Alkaline Earth Cations and Ammonium in Water and Wastewater by Ion Chromatography

ICS
71.040.50 (Physicochemical methods of analysis)
CCS
Z16
发布
2003
实施

1.1 This specification covers requirements for wipe materials that are used to collect settled dusts on hard surfaces for the subsequent determination of lead. 1.2 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

Standard Specification for Wipe Sampling Materials for Lead in Surface Dust

ICS
71.040.50 (Physicochemical methods of analysis)
CCS
Z18
发布
2003
实施

The acquisition of chemical information from variations in the energy position of peaks in the XPS spectrum is of primary interest in the use of XPS as a surface analytical tool. Surface charging acts to shift spectral peaks independent of their chemical relationship to other elements on the same surface. The desire to eliminate the influence of surface charging on the peak positions and peak shapes has resulted in the development of several empirical methods designed to assist in the interpretation of the XPS peak positions, determine surface chemistry, and allow comparison of spectra of conducting and nonconducting systems of the same element. It is assumed that the spectrometer is generally working properly for non-insulating specimens (see Practice E 902). No ideal method has been developed to deal with surface charging (3, 4). For insulators, an appropriate choice of any control or referencing system will depend on the nature of the specimen, the instruments, and the information needed. The appropriate use of charge control and referencing techniques will result in more consistent, reproducible data. Researchers are strongly urged to report both the control and referencing techniques that have been used, the specific peaks and binding energies used as standards (if any), and the criteria applied in determining optimum results so that the appropriate comparisons may be made.1.1 This guide covers the acquainting of the XPS user with the various charge control and charge shift referencing techniques that are and have been used in the acquisition and interpretation of X-ray photoelectron spectroscopy (XPS) data from surfaces of insulating specimens. 1.2 This guide is intended to apply to charge control and charge referencing techniques in XPS and is not necessarily applicable to electron-excited systems. 1.3 SI units are standard unless otherwise noted. 1.4 This standard does not purport to address all of the safety problems, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

Standard Guide to Charge Control and Charge Referencing Techniques in X-Ray Photoelectron Spectroscopy

ICS
71.040.50 (Physicochemical methods of analysis)
CCS
A42
发布
2003
实施

1.1 This guide covers procedures for using a direct current argon plasma atomic emission spectrometer (DCP) to determine the concentration of elements in solution. Recommendations are provided for preparing and calibrating the instrument, assessing instrument performance, diagnosing and correcting for interferences, measuring test solutions, and calculating results. A method to correct for instrument drift is included.1.2 This guide does not specify all the operating conditions for a DCP because of the differences between models of these instruments. Analysts should follow instructions provided by the manufacturer of the particular instrument.1.3 This guide does not attempt to specify in detail all of the hardware components and computer software of the instrument. It is assumed that the instrument, whether commercially available, modified, or custom built, will be capable of performing the analyses for which it is intended, and that the analyst has verified this before performing the analysis.1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use. Specific precautionary statements are given in Section 8.

Standard Guide for Direct Current Plasma Emission Spectrometry Analysis

ICS
71.040.50 (Physicochemical methods of analysis)
CCS
A43
发布
2003
实施

Auger electron spectroscopy yields information concerning the chemical and physical state of a solid surface in the near surface region. Nondestructive depth profiling is limited to this near surface region. Techniques for measuring the crater depths and film thicknesses are given in (35). Ion sputtering is primarily used for depths of less than the order of 1 μm. Angle lapping or mechanical cratering is primarily used for depths greater than the order of 1 μm. The choice of depth profiling methods for investigating an interface depends on surface roughness, interface roughness, and film thickness (1).3 1.1 This guide covers procedures used for depth profiling in Auger electron spectroscopy.1.2 Guidelines are given for depth profiling by the following:SectionIon Sputtering6Angle Lapping and Cross-Sectioning7Mechanical Cratering8Nondestructive Depth Profiling91.3 This standard does not purport to address all of the safety problems, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

Standard Guide for Depth Profiling in Auger Electron Spectroscopy

ICS
71.040.50 (Physicochemical methods of analysis)
CCS
A43
发布
2003
实施

This practice is suitable for the determination of extractable substances that may be found in materials used in systems or components requiring a high level of cleanliness, such as oxygen systems. Soft goods, such as seals and valve seats, may be tested as received. Gloves and wipes, or samples thereof, to be used in cleaning operations may be evaluated prior to use to ensure that the proposed extracting agent does not extract or deposit chemicals, or both, on the surface to be cleaned. Wipes or other cleaning equipment may be tested after use to determine the amount of contaminant removed from a surface. Note 18212;The amount of material extracted may be dependent upon the frequency and power density of the ultrasonic unit. The extraction efficiency has been shown to vary with the frequency and power density of the ultrasonic unit. The unit, therefore, must be carefully evaluated to optimize the extraction conditions.1.1 This practice may be used to extract nonvolatile and semivolatile residues from materials such as new and used gloves, new and used wipes, component soft goods, and so forth. When used with proposed cleaning materials (wipes, gloves, and so forth), this practice may be used to determine the potential of the proposed solvent or other fluids to extract contaminants (plasticizers, residual detergents, brighteners, and so forth.) and deposit them on the surface being cleaned. 1.2 This practice is not suitable for the evaluation of particulate contamination. 1.3 The values stated in SI units are to be regarded standard. No other units of measurement are included in this standard. 1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

Standard Practice for Determination of Soluble Residual Contaminants in Materials by Ultrasonic Extraction

ICS
71.040.50 (Physicochemical methods of analysis)
CCS
G04
发布
2003
实施

The determination of actinides by alpha spectrometry is an essential function of many environmental programs. Alpha spectrometry allows the identification and quantification of most alpha-emitting actinides. Although numerous separation methods are used, the final sample preparation technique has historically been by electrodeposition. However, electrodeposition may have some drawbacks, such as time required, incompatibility with prior chemistry, thick deposits, and low recoveries. These problems can be minimized using the neodymium fluoride method. The sample mounting technique described in this practice is rapid, adds an additional purification step, since only those elements that form insoluble fluorides are mounted, and the sample and filter media can be dissolved and remounted if problems occur. The recoveries are better and resolution approaches normal electrodeposited samples. Recoveries are sufficiently high that for survey work, if quantitative recoveries are not necessary, tracers can be omitted. Drawbacks to this technique include use of very hazardous hydrofluoric acid and the possibility of a non-reproducible and ill-defined counting geometry from filters that are not flat. Also, although the total turn around time for coprecipitation may be less than for electrodeposition, coprecipitation required more time and attention from the analyst.1.1 This practice covers the preparation of separated fractions of actinides for alpha spectrometry as an alternate to electrodeposition. It is applicable to any of the actinides that can be dissolved in dilute hydrochloric acid. Examples of applicable samples would be the final elution from an ion exchange separation or the final strip from a solvent extraction separation.1.2 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use. For a specific hazard statement, see Section 8.

Standard Practice for Mounting Actinides for Alpha Spectrometry Using Neodymium Fluoride

ICS
71.040.50 (Physicochemical methods of analysis)
CCS
H67
发布
2003
实施

The method is applicable to the analysis of materials to demonstrate compliance with the specifications set forth in Specifications C 787 and C 996.1.1 This method covers the determination of plutonium and neptunium isotopes in uranium hexafluoride by alpha spectroscopy. The method could also be applicable to any matrix that may be converted to a nitric acid system.1.2 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory requirements prior to use.

Standard Guide for Determination of Plutonium and Neptunium in Uranium Hexafluoride by Alpha Spectrometry

ICS
71.040.50 (Physicochemical methods of analysis)
CCS
F40
发布
2003
实施

In using Practice D 3960 to measure the regulatory VOC content of coatings, precision tends to be poor for low VOC content waterborne coatings because the VOC weight fraction is determined indirectly. The present method first identifies and then quantifies the weight fraction of individual VOCs directly in low VOC content waterborne air-dry coatings. The total VOC weight fraction can be obtained by adding the individual weight fraction values (Note 3). Note 38212;An effort is currently underway in California to consider changing mass-based VOC regulations for architectural coatings to reactivity-based VOC regulations. In California, reactivity based regulations have already been implemented for aerosol coatings, that is, MIR-indexed regulations (California Air Resources Board). Reactivity based regulations would require knowing the weight fractions of each individual volatile compound present in a coating. SPME/GC makes it possible to identify very low levels of volatile compounds in a coating and could serve to make it possible to identify the presence of hazardous air pollutants (HAPs).1.1 This test method is for the determination of the weight percent of individual volatile organic compounds in low VOC content waterborne latex air-dry coatings. The method is intended primarily for analysis of waterborne coatings in which the material VOC content is below 5 weight percent. The method has been used successfully with higher VOC content waterborne coatings. 1.2 This method may also be used to measure the exempt volatile organic compound content (acetone, methyl acetate, and p-chlorobezotrifluoride) of waterborne coatings. The methodology is virtually identical to that used in Test Method D 6133 and similar to that used in Test Method D 6438. 1.3 Volatile compounds that are present at the 0.05 weight percent level or greater can be determined. Solid phase microextraction will detect volatile compounds at lower levels. 1.4 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard. 1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

Standard Test Method for Speciation of the Volatile Organic Compounds (VOCs) in Low VOC Content Waterborne Air-Dry Coatings by Gas Chromatograpy

ICS
71.040.50 (Physicochemical methods of analysis)
CCS
A29
发布
2003
实施

5.1 Reverse osmosis and nanofiltration desalinating devices can be used to produce potable water from brackish supplies (lt;108201;000 mg/L) and seawater as well as to upgrade the quality of industrial water. These test methods permit the measurement of the performance of reverse osmosis devices using standard sets of conditions and are intended for short-term testing (lt;24 h). These test methods can be used to determine changes that may have occurred in the operating characteristics of reverse osmosis and nanofiltration devices but are not intended to be used for plant design. 1.1 These test methods cover the determination of the operating characteristics of reverse osmosis devices using standard test conditions and are not necessarily applicable to natural waters. Three test methods are given, as follows:   Sections   Test Method A—Brackish Water Reverse Osmosis Devices 8 – 13 Test Method B—Nanofiltration Devices 14 – 19 Test Method B—Seawater Reverse Osmosis Devices 20 – 25 1.2 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

Standard Test Methods for Operating Characteristics of Reverse Osmosis and Nanofiltration Devices

ICS
71.040.50 (Physicochemical methods of analysis)
CCS
Q81
发布
2003
实施

The degree of deacetylation of chitosan salts is an important characterization parameter since the charge density of the chitosan molecule is responsible for potential biological and functional effects. The degree of deacetylation (% DA) of water-soluble chitosan salts can be determined by 1H nuclear magnetic resonance spectroscopy (1H NMR). Several workers have reported on the NMR determination of chemical composition and sequential arrangement of monomer units in chitin and chitosan. The test method described is primarily based on the work of V?/span>rum et al. (1991), which represents the first publication on routine determination of chemical composition in chitosans by solution state 1H NMR spectroscopy. This test method is applicable for determining the % DA of chitosan chloride and chitosan glutamate salts. It is a simple, rapid, and suitable method for routine use. Quantitative 1H NMR spectroscopy reports directly on the relative concentration of chemically distinct protons in the sample, consequently, no assumptions, calibration curves or calculations other than determination of relative signal intensity ratios are necessary. In order to obtain well-resolved NMR spectra, depolymerization of chitosans to a number average degree of polymerization (DPn) of ~15 to 30 is required. This reduces the viscosity and increases the mobility of the molecules. Although there are several options for depolymerization of chitosans, the most convenient procedure is that of nitrous acid degradation in deuterated water. The reaction is selective, stoichiometric with respect to GlcN, rapid, and easily controlled (Allan & Peyron, 1995). The reaction selectively cleaves after a GlcN-residue, transforming it into 2,5-anhydro-D-mannose (chitose), consequently, depletion of GlcN after depolymerization is expected. On the other hand, the chitose unit displays characteristic 1H NMR signals the intensity of which may be estimated and utilized in the calculation of % DA, eliminating the need for correction factors. Using the intensity of the chitose signals, the number average degree of polymerization can easily be calculated as a control of the depolymerization. Samples are equilibrated and analyzed at a temperature of 90 ± 1°C. Elevated sample temperature contributes to reducing sample viscosity and repositions the proton signal of residual water to an area outside that of interest. While samples are not stored at 90°C but only analyzed at this elevated temperature, the NMR tubes should be sealed with a stopper to avoid any evaporation. At a sample pH* of 3.8-4.3 (see 6.1.5 below), artifactual deacetylation of the sample does not occur during the short equilibration and analysis time. A general description of NMR can be found in <761> of the USP24-NF19.1.1 This test method covers the determination of the degree of deacetylation in chitosan and chitosan salts intended for use in biomedical and pharmaceutical applications as well as in Tissue Engineered Medical Products (TEMPs) by high-resolution proton NMR (1H NMR). A guide for the characterization of chitosan salts has been published as Guide F 2103. 1.2 The test method is applicable for determining the degree of deacetylation (% DA) of chitosan chloride and chitosan glutamate salts and is valid for % DA values from 50 up to and including 99. It is simple, rapid, and suitable for routine use. Knowledge of the degree of deacetylation is important for an understanding of the functionality of chitosan salts in TEMP formulations and applications. This test method will assist end users in choosing......

Standard Test Method for Determining Degree of Deacetylation in Chitosan Salts by Proton Nuclear Magnetic Resonance (1H NMR) Spectroscopy

ICS
71.040.50 (Physicochemical methods of analysis)
CCS
C10
发布
2003
实施

This test method is suitable for setting specifications on phenol and for use as an internal quality control tool where phenol is produced or is used in a manufacturing process. It may also be used in development or research work involving phenol. It is generally applied to determining those commonly occurring impurities such as mesityl oxide, cumene, hydroxyacetone, acetone, alpha-methylstyrene, 2-methylbenzofuran, acetophenone, o-cresol, and m & p-cresol. Purity is commonly reported by subtracting the determined expected impurities from 100.00. However, a gas chromatographic analysis cannot determine absolute purity if unknown components are contained within the material being examined.1.1 This test method covers the determination of known impurities in phenol by gas chromatography (GC). It is generally meant for the analysis of phenol of 99.9% or greater purity. 1.2 The following applies to all specified limits in this test method: for purposes of determining conformance with this test method, an observed value or calculated value shall be rounded off "to the nearest unit" in the last right-hand digit used in expressing the specification limit in accordance with the rounding-off method of Practice E 29. Precision data is based on impurity concentrations of 15 to 70 mg/kg. Users of this test method believe it is linear over a wider range. 1.3 This Test Method does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use. For specific hazard statements, see Section 9.

Standard Test Method for Analysis of Phenol by Capillary Gas Chromatography

ICS
71.040.50 (Physicochemical methods of analysis)
CCS
G17
发布
2003
实施

1.1 This practice describes the determination of the relative actinic effects of individual spectral bands of an exposure source on a material. The activation spectrum is specific to the light source to which the material is exposed to obtain the activation spectrum. A light source with a different spectral power distribution will produce a different activation spectrum. 1.2 This practice describes two procedures for determining an activation spectrum. One uses sharp cut-on UV/visible transmitting filters and the other uses a spectrograph to determine the relative degradation caused by individual spectral regions. Note 1-Other techniques can be used to isolate the effects of individual spectral bands of a light source, for example, interference filters. 1.3 The techniques are applicable to determination of the spectral effects of solar radiation and laboratory accelerated test devices on a material. They are described for the UV region, but can be extended into the visible region using different cut-on filters and appropriate spectrographs.1.4 The techniques are applicable to a variety of materials, both transparent and opaque, including plastics, paints, inks, textiles and others. 1.5 The optical and/or physical property changes in a material can be determined by various appropriate methods. The methods of evaluation are beyond the scope of this practice. 1.6 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.Note 2-There is no ISO standard that is equivalent to this standard.

Standard Practice for Determining the Activation Spectrum of a Material (Wavelength Sensitivity to an Exposure Source) Using the Sharp Cut-On Filter or Spectrographic Technique

ICS
71.040.50 (Physicochemical methods of analysis)
CCS
G04
发布
2003
实施

Reverse osmosis and nanofiltration desalinating devices can be used to produce potable water from brackish supplies (<10 000 mg/L) and seawater as well as to upgrade the quality of industrial water. These test methods permit the measurement of the performance of reverse osmosis devices using standard sets of conditions and are intended for short-term testing (<24 h). These test methods can be used to determine changes that may have occurred in the operating characteristics of reverse osmosis and nanofiltration devices but are not intended to be used for plant design.1.1 These test methods cover the determination of the operating characteristics of reverse osmosis devices using standard test conditions and are not necessarily applicable to natural waters. Three test methods are given, as follows:SectionsTest Method A8212;Brackish Water Reverse Osmosis Devices8-13Test Method B8212;Nanofiltration Devices14-19Test Method C8212;Seawater Reverse Osmosis Devices20-251.2 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

Standard Test Methods for Operating Characteristics of Reverse Osmosis and Nanofiltration Devices

ICS
71.040.50 (Physicochemical methods of analysis)
CCS
J77
发布
2003
实施

In using Practice D 3960 to measure the regulatory VOC content of coatings, precision tends to be poor for low VOC content waterborne coatings because the VOC weight fraction is determined indirectly. The present method first identifies and then quantifies the weight fraction of individual VOCs directly in low VOC content waterborne air-dry coatings. The total VOC weight fraction can be obtained by adding the individual weight fraction values (Note 3). Note 38212;An effort is currently underway in California to consider changing mass-based VOC regulations for architectural coatings to reactivity-based VOC regulations. In California, reactivity based regulations have already been implemented for aerosol coatings, that is, MIR-indexed regulations (California Air Resources Board). Reactivity based regulations would require knowing the weight fractions of each individual volatile compound present in a coating. SPME/GC makes it possible to identify very low levels of volatile compounds in a coating and could serve to make it possible to identify the presence of hazardous air pollutants (HAPs).1.1 This test method is for the determination of the weight percent of individual volatile organic compounds in low VOC content waterborne latex air-dry coatings. The method is intended primarily for analysis of waterborne coatings in which the material VOC content is below 5 weight percent. The method has been used successfully with higher VOC content waterborne coatings.1.2 This method may also be used to measure the exempt volatile organic compound content (acetone, methyl acetate, and p-chlorobezotrifluoride) of waterborne coatings. The methodology is virtually identical to that used in Test Method D 6133 and similar to that used in Test Method D 6438. 1.3 Volatile compounds that are present at the 0.05 weight percent level or greater can be determined. Solid phase microextraction will detect volatile compounds at lower levels.1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

Standard Test Method for Speciation of the Volatile Organic Compounds (VOCs) in Low VOC Content Waterborne Air-Dry Coatings by Gas Chromatograpy

ICS
71.040.50 (Physicochemical methods of analysis)
CCS
发布
2003
实施

This test method calibrates or demonstrates conformity of thermogravimetric apparatus at ambient conditions. Most thermogravimetry analysis experiments are carried out under temperature ramp conditions or at isothermal temperatures distant from ambient conditions. This test method does not address the temperature effects on mass calibration. In most thermogravimetry experiments, the mass change is reported as weight percent in which the observed mass at any time during the course of the experiment is divided by the original mass of the test specimen. This method of reporting results assumes that the mass scale of the apparatus is linear with increasing mass. In such cases, it may be necessary only to confirm the performance of the instrument by comparison to a suitable reference. When the actual mass of the test specimen is recorded, the use of a calibration factor to correct the calibration of the apparatus may be required, on rare occasions.1.1 This test method covers the calibration or performance confirmation of the mass (or weight) scale of thermogravimetric analyzers and is applicable to commercial and custom-built apparatus. 1.2 Electronic instrumentation or automated data analysis and reduction systems or treatments equivalent to this test method may be used. 1.3 The values stated in SI units are to be regarded as the standard.

Standard Test Method for Mass Scale Calibration of Thermogravimetric Analyzers

ICS
71.040.50 (Physicochemical methods of analysis)
CCS
N11
发布
2003
实施

1.1 This specification covers requirements for wipes that are used to collect settled dusts on surfaces for the subsequent determination of lead. 1.2 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard. 1.3 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

Standard Specification for Wipe Sampling Materials for Lead in Surface Dust

ICS
71.040.50 (Physicochemical methods of analysis)
CCS
Z18
发布
2003
实施

1.1 This test method is for the determination of gold in copper concentrates in the concentration range from 0.2 to 17 956;g/g (0.007 to 0.500 Troy oz/short ton). Note 18212;The lower scope limit is set in accordance with Practice E 1601.1.2 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use. For specific warning statements, see 11.3.1, 11.5.4, and 11.6.5.

Standard Test Method for Fire Assay Determination of Gold in Copper Concentrates by Gravimetry

ICS
71.040.50 (Physicochemical methods of analysis)
CCS
发布
2002
实施

1.1 This test method covers a procedure for the determination of the transmittance of monoethylene glycol (1,2-ethanediol; MEG) at wavelengths in the region 220 to 350 nm. The results provide a measure of the purity of the sample with respect to ultraviolet absorbing compounds.1.2 The values stated in SI units are to be regarded as the standard.1.3 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.1.3 Review the current Material Safety Data Sheets (MSDS) for detailed information concerning toxicity, first aid procedures, and safety precautions.

Standard Test Method for Ultraviolet Transmittance of Monoethylene Glycol (Ultraviolet Spectrophotometric Method)

ICS
71.040.50 (Physicochemical methods of analysis)
CCS
G17
发布
2002
实施

1.1 This test method covers the determination of silver in the range of 50 g/g to 1000 g/g by acid dissolution of the silver and measurement by atomic absorption spectrophotometry. Copper concentrates are internationally traded within the following concentration ranges:ElementUnit Concentration RangeAluminum%0.05to 2.50Antimony%0.0001to 4.50Arsenic%0.01to 0.50Barium%0.003to 0.10Bismuth%0.001to 0.16Cadmium%0.0005to 0.04Calcium%0.05to 4.00Carbon %0.10to 0.90Chlorine%0.001to 0.006Chromium%0.0001to 0.10Cobalt%0.0005to 0.20Copper%10.0to 44.0Fluorine%0.001to 0.10Gold g/g1.40to100.0Iron %12.0to30.0Lead %0.01to1.40Magnesium%0.02to 2.00Manganese%0.009to 0.10Mercuryg/g0.05to 50.0Molybdenum% 0.002to 0.25Nickel%0.0001to 0.08Silicon%0.40to 20.0Silverg/g18.0to 8000Sulfur%10.0to 36.0Tin %0.004to0.012Zinc %0.005to4.301.2 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to its use./p>

Standard Test Method for Determination of Silver in Copper Concentrates by Flame Atomic Absorption Spectrometry

ICS
71.040.50 (Physicochemical methods of analysis)
CCS
D42
发布
2002
实施



Copyright ©2007-2022 ANTPEDIA, All Rights Reserved
京ICP备07018254号 京公网安备1101085018 电信与信息服务业务经营许可证:京ICP证110310号