ELISA: Direct

2020.9.21

实验概要

The following protocol provides a method of direct ELISA by Peprotech.

主要试剂

Tween-20 (Sigma Cat. # P-7949)
BSA (Sigma Cat. # A-7030) 
ABTS Liquid Substrate Solution (Sigma Cat. # A3219) 
Dulbecco’s PBS [10x] (Gibco BRL Cat. #14200-075)
 
All solutions should be at ambient temperature prior to use.
PBS: Dilute 10xPBS to 1xPBS, pH 7.20 in sterile water.
Wash Buffer: 0.05% Tween-20 in PBS
Block Buffer: 1% BSA in PBS*
Diluent: 0.05% Tween-20, 0.1% BSA in PBS*
Substrate Solution: ABTS Liquid Substrate Solution
*Sterile filter and store at 4°C for up to 1 week.

主要设备

ELISA microplates (Nunc MaxiSorp Prod. # 439454 or Corning Prod. # 3590)
Avidin-HRP conjugate (Sigma Cat. # A-7419)

实验材料

PeproTech’s Recombinant Protein
PeproTech’s Biotinylated Antigen Affinity Purified Polyclonal Antibody
**PeproTech’s Standards and Antibodies should be reconstituted according to the data sheets which accompany each product.

实验步骤

1. Plate preparation

    1) Standard/Sample: Serial dilute standard from 0.1μg/ml to zero  in PBS. Add 100μl of standard or sample to each well in triplicate.  Incubate at room temperature overnight.

    2) Aspirate the wells to remove liquid and wash plates 4 times.  Each wash consists of adding 300μl wash buffer per well, followed by  aspiration. After the last wash invert plate to remove residual buffer  and blot on paper towel.

    3) Add 300μl blocking buffer to each well. Incubate 2 hour at R.T.

    4) Aspirate and wash plate 4 times (as in step 2).

2. ELISA protocol
    1) Detection:Wash  plate four times. Dilute detection antibody (biotinylated) in diluent  to a concentration of 1 μg/ml. Immediately add 100μl per well. Incubate  at room temperature for 2 hours.
    2) Avidin-HRP Conjugate:Aspirate and wash plate 4 times. Dilute Avidin-HRP conjugate 1:2000 in diluent.
Add 100μl per well. Incubate 30 min at room temperature.
     3) Substrate: Aspirate  and wash plate 4 times. Add 100μl of substrate solution to each well.  Incubate at room temperature for color development. Monitor color  development with an ELISA plate reader at 405 nm with wavelength  correction set at 650 nm.

注意事项

1. Reliable  standard curves are obtained when O.D. readings do not exceed 0.2 units  for the zero standard concentrations, or 1.2 units for the highest  standard concentration. The plate should be monitored at 5 minute  intervals until desired O.D. readings are obtained. The typical range is  5-40 minutes. O.D. readings may vary.
2. Assay sensitivity may be increased with additional washings.