Analysis and Validation of a Psychrometric Apparatus


 

 

非常抱歉,我们暂时无法提供预览,您可以试试: 免费下载 4704-2004 前三页,或者稍后再访问。

如果您需要购买此标准的全文,请联系:

点击下载后,生成下载文件时间比较长,请耐心等待......

 

标准号
4704-2004
发布日期
2004年01月01日
实施日期
2005年01月20日
废止日期
中国标准分类号
/
国际标准分类号
/
发布单位
ASHRAE - American Society of Heating@ Refrigerating and Air-Conditioning Engineers@ Inc.
引用标准
8
适用范围
This article uses the analysis ofvariance as a tool to validate a calibration apparatus for temperature and relative humidity sensors. The validation procedure establishes the operational ranges for measurement of the dry-bulb (tdb) and wet-bdb (twb) temperatures andofthe relative humidity of air (RH) for this particular apparatus. Analyses are carried out testing twb as afunction of air speed ranging from I . 25 to 5.00 m/s and twb as afunction of RHand also compares two environments to calibrate tdb: either in thermostatic bath or in the airjlow of a test section. The results show that air speed on wet covered semors within the tested range of I . 25 io 5. O m/s does not change the measured twb@ thus widening the range suggested in ASHRAEStandard41.1-1974.H owevecfor relative humidity of air within the studied range@ significant differences are found@ and these are close to the 2-3% indicated range. The use of different calibrated sensorsfor tdb result in different calculated RHc values in a controlled environment.