XR 28-2013

Supplemental Requirements for User Information and System Function Related to Dose in CT


 

 

非常抱歉,我们暂时无法提供预览,您可以试试: 免费下载 XR 28-2013 前三页,或者稍后再访问。

如果您需要购买此标准的全文,请联系:

点击下载后,生成下载文件时间比较长,请耐心等待......

 

标准号
XR 28-2013
发布日期
2013年01月01日
实施日期
2015年05月13日
废止日期
中国标准分类号
/
国际标准分类号
/
发布单位
NEMA - National Electrical Manufacturers Association
引用标准
27
适用范围
"This standard identifies uniform and standardized manufacturer's information provided to users of a CT scanner. Users include people designated through an Access Control system to operate a CT scanner according to the manufacturer's user instructions@ to authorize protocols in the facility where the scanner is housed@ and to perform recommended quality control testing on the CT scanner. This user information pertains to perfusion scanning@ use of Automatic Exposure Control (AEC)@ organization of dose-related information@ and a requirement for listing reference protocols shipped on the CT system. This user information also requires pre-populating the NEMA XR-25 Dose Check ""Alert"" value at a CTDIvol value not to exceed 1 Gy (1000 mGy)@ as well as system function when switching between a manual mA control mode and automatic exposure control. This standard is established to collect the standardized responses and resolutions to FDA's November 8@ 2010 ""Letter to the Medical Imaging Technology Alliance Regarding CT Recommendations."" (See Appendix D.) This standard provides supplementary requirements to those already in place for information provided to the user from other normative documents (see Section 1.3@ references)."