5.1 This practice can be used to determine if a constant, proportional, or linear bias correction can improve the degree of agreement between two methods that purport to measure the same property of a material.
5.2 The bias correction developed in this practice can be applied to a single result (X) obtained from one test method (method X) to obtain a predicted result ( Y^) for the other test method (method Y).
Note 6: Users are cautioned to ensure that Y^ is within the scope of method Y before its use.
5.3 The between methods reproducibility established by this practice can be used to construct an interval around Y^ that would contain the result of test method Y, if it were conducted, with about 958201;% confidence.
5.4 This practice can be used to guide commercial agreements and product disposition decisions involving test methods that have been evaluated relative to each other in accordance with this practice.
5.5 The magnitude of a statistically detectable bias is directly related to the uncertainties of the statistics from the experimental study. These uncertainties are related to both the size of the data set and the precision of the processes being studied. A large data set, or, highly precise test method(s), or both, can reduce the uncertainties of experimental statistics to the point where the “statistically detectable” bias can become “trivially small,” or be considered of no practical consequence in the intended use of the test method under study. Therefore, users of this practice are advised to determine in advance as to the magnitude of bias correction below which they would consider it to be unnecessary, or, of no practical concern for the intended application prior to execution of this practice.
Note 7: It should be noted that the determination of this minimum bias of no practical concern is not a statistical decision, but rather, a subjective decision that is directly dependent on the application require...... 这两种原料都用于香水和化妆品行业。为了显示试验的分散性和重复性,将表2所列精油和香料的闪点结果分为三份进行测定。结果如表3所示,列出了一系列“无闪点”的结果,主要是因为手霜、身体霜、化妆品行业使用的其他添加剂和原材料的含水量相当大。表3还比较了两种不同类型的测量结果:快速扫描(SCAN)和ASTM D7094。根据ASTM D7094测量闪点,需要知道预期闪点。... 结果根据ASTM D7545和EN 16091标准,使用浓度增加的点火改进剂对两种不同的柴油样品(纯柴油和随着浓度增加的硝酸2-乙基己酯(EHN))。相应的诱导期如表2所示。正如预期的,结果表明,当EHN浓度增加时,两种柴油燃料样品的氧化稳定性降低(参见图3)。...用安东帕RapidOxy 100 Fuel进行柴油试验
Copyright ©2007-2022 ANTPEDIA, All Rights Reserved