F81 通用核仪器 标准查询与下载



共找到 220 条与 通用核仪器 相关的标准,共 15

Based on the recommendations of the major emitter couples logic (ECL) manufacturers for communication between different parts of a system. They advise differential line driving and receiving for high noise immunity and cancellation of ground potential di

Nuclear instrumentation - ECL (Emitter Coupled Logic) front panel interconnections in counter logic

ICS
27.120.01;31.220.10
CCS
F81
发布
1996-06
实施

The purpose of this International Standard is to provide the user with a means of verifying the performance of typical liquid-scintillation counting systems. Measures of performance considered in this standard are: - counting system efficiency; - reproducibility of sample and background count rates. This standard does not cover the calculation of sample activity for quenched samples. Accordingly, this standard does not deal with sample preparation, efficiency correlation (quench correction) procedures, or the identification of unknown radionuclides.

Nuclear instrumentation. Liquid-scintillation counting systems. Performance verification

ICS
27.120.10
CCS
F81
发布
1995-02-15
实施
1995-02-15

Radiophotoluminescent glass dosemeter systems

ICS
17.240
CCS
F81
发布
1995-02-01
实施

This International Standard is applicable to multichannel pulse height (amplitude) analyzers (MCA) with linear amplitude response, which are measuring devices used for acquisition, storage and processing of amplitude distributions. These analyzers automatically carry out the following operations: - acceptance of pulses from detection sub-assemblies or other sources; - analogue-to-digital conversion (ADC) of the pulse height information; - sorting of pulse height data according to predetermined characteristics; - storage of digital information (histogram); - processing of stored pulse height histogram and externally supplied information in accordance with predetermined algorithms; - data input and output functions (for example, driving a display, a printer, a floppy disk drive, a plotter, etc.). The above operations may be accomplished separately or combined in an integrated instrument. All the tests described herein are not mandatory, but if the test is carried out, then it shall be done in accordance with the procedures given herein. In many cases in nuclear physics, it is very important to measure the distribution of certain kinds of parameters, such as the energy of particles, their mass, their distribution in time, their scattering on certain angles, etc. In modern measuring practice, this performance is usually accomplished by means of a multichannel pulse height analyzer. This type of a device is built in such a way that a quantity of a certain physical quality is digitized and then stored according to its digitized number. All modern MCAs contain special storing units (memories) for this purpose. After accomplishing the measurement the required information is retrieved from the memory so that it can be used for interpretation of the physical data. Nowadays, multichannel pulse height analyzers are more widely used in different fields of science and industry. Their specific aim in nuclear applications is to transform the pulse height related to a certain nuclear event (usually energy) into a number corresponding to a memory cell, so that during the measurement one can store a histogram that is very similar to the initial pulse height distribution. This distribution reflects the probability density of particle energy (energy spectrum). Using the stored information about the distribution it is easy to determine the information about the initial energy of the particles (alpha and/or beta-particles, gamma and X-quantum, etc.). Further, this information can be used to determine flux densities, dose rates and doses, isotope and element concentrations and contents, etc. Usually, a pulse height analyzer includes an ADC (analogue-to-digital converter), an acquisition unit, a memory, a display and different input/output (I/O) units (display, printer, plotter, etc.). The object of this standard is to present terms and definitions, to list main characteristics of multichannel analyzers, technical requirements and test methods for the following parameters: - minimum and maximum signal pulse heights to be measured; · main error of the minimum and maximum measured signal pulse heights; · instability of the minimum and maximum measured signal pulse heights; · additional errors (or variations) of the minimum and maximum measured signal pulse heights; - channel width; · total error of the channel width; · instability of the channel width; · additional errors of the channel width; - zero point position; · offset; · instability of the zero point; · additional errors of the zero point; - operating range; - integral non-linearity; - differential non-linearity; - dead time; - maximum input pulse rate to be measured; - live time error; - non rectangular factor of channel profile; - system throughput.

Nuclear instrumentation - Multichannel pulse height analyzers - Main characteristics, technical requirements and test methods

ICS
27.120.10
CCS
F81
发布
1995-02
实施

Radiation protection instrumentation - Direct reading personal dose equivalent (rate) monitors - X, gamma and high energy beta radiation

ICS
CCS
F81
发布
1995-02
实施
2005-02-16

Radiation protection instrumentation - Neutron radiation - Direct reading personal dose equivalent and/or dose equivalent rate monitors

ICS
CCS
F81
发布
1995-01
实施
2005-02-16

本部分适用于核谱测量中脉冲幅度直方图数据的交换,而与数据来源、读或写数据设备以及容纳数 据的媒体无关。 本部分的目的是提供核谱测量用多道分析器(MCA)直方图数据的交换格式,该交换格式能用于各 实验室间传送多道脉冲幅度数据以及为检验目的而分配这些数据。为与大量的计算机语言、计算机以 及硬件链路相兼容,必须用ASCH书写完整的文件。预期这些文件在使用前将转换为本地格式。

Nuclear instrumentation - MCA histogram data interchange format for nuclear spectroscopy

ICS
27.120.10
CCS
F81
发布
1995-01
实施

The purpose of this standard is to specify the design requirements and performance characteristics of personal warning devices used to give an audible or audible and visual indications related to dose equivalent rate from strongly penetrating raditions(a

Radiation Protection Instrumentation - Monitoring Equipment - Personal Warning Devices for X and Gamma Radiations

ICS
13.280
CCS
F81
发布
1995
实施

1.1 This test method covers the calibration of ultraviolet light-measuring radiometers possessing either narrow- or broad-band spectral response distributions using either a scanning or a linear-diode-array spectroradiometer as the primary reference instrument. For transfer of calibration from radiometers calibrated by this test method to other instruments, Test Method E824 should be used. Note 1-Special precautions must be taken when a diode-array spectroradiometer is employed in the calibration of filter radiometers having spectral response distributions below 320-nm wavelength. Such precautions are described in detail in subsequent sections of this test method. 1.2 This test method is limited to calibrations of radiometers against light sources that the radiometers will be used to measure during field use. Note 2-For example, an ultraviolet radiometer calibrated against natural sunlight cannot be employed to measure the total ultraviolet irradiance of a fluorescent ultraviolet lamp. 1.3 Calibrations performed using this test method may be against natural sunlight, Xenon-arc burners, metal halide burners, tungsten and tungsten-halogen lamps, fluorescent lamps, etc. 1.4 Radiometers that may be calibrated by this test method include narrow-, broad-, and wide-band ultraviolet radiometers, and narrow-, broad, and wide-band visible-region-only radiometers, or radiometers having wavelength response distributions that fall into both the ultraviolet and visible regions. Note 3-For purposes of this test method, narrow-band radiometers are those with [delta][lambda] [and;lt;=] 20 nm, broad-band radiometers are those with 20 nm [and;lt;=] [delta][lambda] [and;lt;=] 70 nm, and wide-band radiometers are those with [delta][lambda] [>=] 70 nm. Note 4-For purposes of this test method, the ultraviolet region is defined as the region from 285 to 400-nm wavelength, and the visible region is defined as the region from 400 to 750-nm wavelength. The ultraviolet region is further defined as being either UV-A with radiation of wavelengths from 315 to 400 nm, or UV-B with radiation from 285 to 315-nm wavelength. 1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

Standard Test Method for Calibration of Narrow- and Broad-Band Ultraviolet Radiometers Using a Spectroradiometer

ICS
CCS
F81
发布
1995
实施

1.1 This test method covers the calibration of ultraviolet light-measuring radiometers possessing either narrow- or broad-band spectral response distributions using either a scanning or a linear-diode-array spectroradiometer as the primary reference instrument. For transfer of calibration from radiometers calibrated by this test method to other instruments, Test Method E 824 should be used.Note 18212;Special precautions must be taken when a diode-array spectroradiometer is employed in the calibration of filter radiometers having spectral response distributions below 320-nm wavelength. Such precautions are described in detail in subsequent sections of this test method.1.2 This test method is limited to calibrations of radiometers against light sources that the radiometers will be used to measure during field use.Note 28212;For example, an ultraviolet radiometer calibrated against natural sunlight cannot be employed to measure the total ultraviolet irradiance of a fluorescent ultraviolet lamp.1.3 Calibrations performed using this test method may be against natural sunlight, Xenon-arc burners, metal halide burners, tungsten and tungsten-halogen lamps, fluorescent lamps, etc.1.4 Radiometers that may be calibrated by this test method include narrow-, broad-, and wide-band ultraviolet radiometers, and narrow-, broad, and wide-band visible-region-only radiometers, or radiometers having wavelength response distributions that fall into both the ultraviolet and visible regions.Note 38212;For purposes of this test method, narrow-band radiometers are those with 916;955; 8804; 20 nm, broad-band radiometers are those with 20 nm 8804;916;955; 8804; 70 nm, and wide-band radiometers are those with 916;955; 8805; 70 nm.Note 48212;For purposes of this test method, the ultraviolet region is defined as the region from 285 to 400-nm wavelength, and the visible region is defined as the region from 400 to 750-nm wavelength. The ultraviolet region is further defined as being either UV-A with radiation of wavelengths from 315 to 400 nm, or UV-B with radiation from 285 to 315-nm wavelength.1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

Standard Test Method for Calibration of Narrow- and Broad-Band Ultraviolet Radiometers Using a Spectroradiometer

ICS
17.180.20 (Colours and measurement of light)
CCS
F81
发布
1995
实施

Scintillator Performance Measurement Method Temperature Effect

ICS
CCS
F81
发布
1994-10-24
实施
1995-01-01

Scintillator Performance Measurement Method Scintillation Decay Time

ICS
CCS
F81
发布
1994-10-24
实施
1995-01-01

Scintillator Performance Measurement Method Emission Spectroscopy

ICS
CCS
F81
发布
1994-10-24
实施
1995-01-01

Scintillator performance measurement method scintillation effective attenuation length

ICS
CCS
F81
发布
1994-10-24
实施
1995-01-01

Intrinsic Amplitude Resolution of Scintillator Performance Measurement Methods

ICS
CCS
F81
发布
1994-10-24
实施
1995-01-01

Scintillator Performance Measurement Method Relative Energy Conversion Efficiency

ICS
CCS
F81
发布
1994-10-24
实施
1995-01-01

Scintillator Performance Measurement Methods Light Output

ICS
CCS
F81
发布
1994-10-24
实施
1995-01-01

General Rules for Measuring Methods of Scintillator Properties

ICS
CCS
F81
发布
1994-10-24
实施
1995-01-01

Gamma directional radiation instrument

ICS
17.240
CCS
F81
发布
1994-07-18
实施
1994-12-01

scaler

ICS
17.240
CCS
F81
发布
1994-03-07
实施
1994-08-01



Copyright ©2007-2022 ANTPEDIA, All Rights Reserved
京ICP备07018254号 京公网安备1101085018 电信与信息服务业务经营许可证:京ICP证110310号