DIN EN ISO 11138-2-2009 保健产品的灭菌.生物指示剂.第2部分:环氧乙烷灭菌过程的生物指示器(ISO 11138-2-2006).英文版本DIN EN ISO 11138-2-2009-09
Sterilization of health care products - Biological indicators - Part 2: Biological indicators for ethylene oxide sterilization processes (ISO 11138-2:2006); English version of DIN EN ISO 11138-2:2009-09
This part of ISO 11138 provides specific requirements for test organisms, suspensions, inoculated carriers,
biological indicators and test methods intended for use in assessing the performance of sterilizers and
sterilization processes employing ethylene oxide gas as the sterilizing agent, either as pure ethylene oxide
gas or mixtures of this gas with diluent gases, at sterilizing temperatures within the range of 29℃ to 65℃.
NOTE 1 Requirements for validation and control of ethylene oxide sterilization processes are provided by ISO 11135.
NOTE 2 National or regional regulations could provide requirements for work place safety.