This standard specifies the protective performance of medical X-ray computed tomography equipment (abbreviated as CT), the protective facilities of the computer room and the radiation protection requirements and testing requirements for safe operation. This standard applies to the protection of CT in use.
GBZ 165-2012 Referenced Document
GB 16348 Radiological protection standards for the examinee in medical X-ray diagnosis
GB 17589 Specifications for quality assurance test for computed tomography X-ray scanners
GB 9706.11 Medical electrical equipment. Part 2: Particular requirements for the safety of X-ray source assemblies and X-ray tube assemblies for medical diagnosis
GB 9706.12 Medical electrical equipment. Part 1: General requirements for safety 3. Collateral standard: General requirements for radiation protection in diagnostic X-ray equipment
GBZ 179 Basic principles for radiological protection of medical exposure
GBZ 165-2012 history
2012GBZ 165-2012 Radiological protection requirements for X-ray computed tomography
2006GBZ 165-2005 Radiological Protection Standards for X-ray Computed Tomography