N34 光学计量仪器 标准查询与下载



共找到 176 条与 光学计量仪器 相关的标准,共 12

Although it is possible to observe and measure each of the several characteristics of a detector under different and unique conditions, it is the intent of this practice that a complete set of detector specifications should be obtained under the same operating conditions. It should also be noted that to completely specify a detector's capability, its performance should be measured at several sets of conditions within the useful range of the detector. The terms and tests described in this practice are sufficiently general that they may be used regardless of the ultimate operating parameters. Linearity and response time of the recorder or other readout device used should be such that they do not distort or otherwise interfere with the performance of the detector. This requires adjusting the gain, damping, and calibration in accordance with the manufacturer's directions. If additional electronic filters or amplifiers are used between the detector and the final readout device, their characteristics should also first be established. 1.1 This practice covers the testing of the performance of a variable-wavelength photometric detector (VWPD) used as the detection component of a liquid-chromatographic (LC) system operating at one or more wavelengths in the range 190 to 800 nm. Many of the measurements are made at 254 nm for consistency with Practice E685. Measurements at other wavelengths are optional. 1.2 This practice is intended to describe the performance of the detector both independently of the chromatographic system (static conditions) and with flowing solvent (dynamic conditions). 1.3 For general liquid chromatographic procedures, consult Refs (1-9). 1.4 For general information concerning the principles, construction, operation, and evaluation of liquid-chromatography detectors, see Refs (10, 11) in addition to the sections devoted to detectors in Refs (1-7). 1.5 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard. 1.6 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

Standard Practice for Testing Variable-Wavelength Photometric Detectors Used in Liquid Chromatography

ICS
71.040.50 (Physicochemical methods of analysis)
CCS
N34
发布
1998
实施

Direct-reading polychromators are instruments commonly used for multi-element spectrochemical analysis. This guide seeks to describe those aspects of such instruments that are of significance in achieving useful spectrochemical performance. Awareness of parameters described in this practice will make manufacturers cognizant of factors they should consider in designing instruments, assist purchasers of instruments in making intelligent comparisons of competing designs, and make users aware of the compromises they must make in performing particular determinations. Adequate description of spectrometers permits forming qualified appraisals on three important performance characteristics: accuracy of analysis, detection limits, and freedom from line interferences.1.1 This guide covers features of a spectrometer or polychromator used for optical emission, direct-reading, spectrochemical analysis. A polychromator in this sense consists of a spectrometer with an extended and fixed wavelength range and an array of fixed exit slits to isolate the spectral lines of the elements to be measured. 1.1.1 This guide does not apply to direct-reading systems that employ echelle spectrometers and vidicon or other detectors where the design parameters are quite different.1.2 This guide covers only the optical portion of the instrument, from excitation stand to photomultipliers.1.2.1 Only general statements are made about source units.1.2.2 Photomultipliers are included to the extent that they are mounted within the spectrometer to convert optical intensities to electrical signals, and establish the instrumental precision of each channel as a light measuring device. Readout systems are not included.1.3 It is not the purpose of this guide to establish binding specifications or tolerances, but rather, to call attention to important parameters that manufacturers should include in their literature, to provide methods for measuring those parameters, and to assign values that are indicative of acceptably good performance. Because of the great variety of demands imposed by spectrochemical techniques, rigid performance criteria are not feasible.1.4 The values stated in SI units are to be regarded as the standard.1.5 This standard does not purport to address all of the safety problems, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

Standard Guide for Describing and Specifying the Spectrometer of an Optical Emission Direct-Reading Instrument

ICS
17.180.30
CCS
N34
发布
1998
实施

Although it is possible to observe and measure each of the several characteristics of a detector under different and unique conditions, it is the intent of this practice that a complete set of detector specifications should be obtained under the same operating conditions. It should also be noted that to completely specify a detectorrsquo; capability, its performance should be measured at several sets of conditions within the useful range of the detector. The terms and tests described in this practice are sufficiently general that they may be used regardless of the ultimate operating parameters. Linearity and response time of the recorder or other readout device used should be such that they do not distort or otherwise interfere with the performance of the detector. This requires adjusting the gain, damping, and calibration in accordance with the manufacturerrsquo;directions. If additional electronic filters or amplifiers are used between the detector and the final readout device, their characteristics should also first be established. 1.1 This practice covers the testing of the performance of a variable-wavelength photometric detector (VWPD) used as the detection component of a liquid-chromatographic (LC) system operating at one or more wavelengths in the range 190 to 800 nm. Many of the measurements are made at 254 nm for consistency with Practice E 685. Measurements at other wavelengths are optional.1.2 This practice is intended to describe the performance of the detector both independently of the chromatographic system (static conditions) and with flowing solvent (dynamic conditions).1.3 For general liquid chromatographic procedures, consult Refs (1-9).1.4 For general information concerning the principles, construction, operation, and evaluation of liquid-chromatography detectors, see Refs (10, 11)in addition to the sections devoted to detectors in Refs (1-7).1.5 The values stated in SI units are to be regarded as standard.1.6 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

Standard Practice for Testing Variable-Wavelength Photometric Detectors Used in Liquid Chromatography

ICS
71.040.50 (Physicochemical methods of analysis)
CCS
N34
发布
1998
实施

この規格は,液体中の粒子の大きさ及び数を測定する光散乱式自動粒子計数器(以下,粒子計数器という。)について規定する。

Light scattering automatic particle counter for liquid

ICS
19.120
CCS
N34
发布
1997-09-20
实施

この規格は,連続的に空気を吸引し,その中の浮遊微粒子の粒径及び個数を測定し,空気中の粒子濃度を求める光散乱式自動粒子計数器(以下,粒子計数器という。)について規定する。

Light scattering automatic particle counter

ICS
19.120
CCS
N34
发布
1997-09-20
实施

General technical requirements for atomic fluorescence photometer

ICS
17.180.30
CCS
N34
发布
1997-07-01
实施
1998-01-15

TCY low-light objective lens T-number measuring instrument verification regulations

ICS
CCS
N34
发布
1997
实施

Test method for maximum range extinction ratio of solid-state pulsed laser range finder

ICS
CCS
N34
发布
1997
实施

This International Standard specifies requirements for continuously indicating focimeters and digitally rounding focimeters with which the vertex powers and prismatic powers of spherical and astigmatic lenses, including lenses mounted in frames and contact lenses, can be measured and with which lenses can be orientated and marked. NOTE 1 For the measurement of vertex powers of contact lenses, see ISO 9337:—, Optics and optical instruments — Contact lenses — Determination of back vertex power. note: 1) To be published.

Optics and optical instruments - Focimeters

ICS
11.040.70
CCS
N34
发布
1996-08-01
实施

1.1 The purpose of this guide is to familiarize the analyst with some of the relevant literature describing the physical properties of modern electrostatic electron spectrometers. 1.2 this guide is intended to apply to electron spectrometers generally used in Auger electron spectroscopy (AES) and X-ray photoelectron spectroscopy (XPS). 1.3 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

Standard Guide for Literature Describing Properties of Electrostatic Electron Spectrometers

ICS
CCS
N34
发布
1996
实施

1.1 The purpose of this guide is to familiarize the analyst with some of the relevant literature describing the physical properties of modern electrostatic electron spectrometers. 1.2 this guide is intended to apply to electron spectrometers generally used in Auger electron spectroscopy (AES) and X-ray photoelectron spectroscopy (XPS). 1.3 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

Standard Guide for Literature Describing Properties of Electrostatic Electron Spectrometers

ICS
17.220.20 (Measurement of electrical and magnetic
CCS
N34
发布
1996
实施

1.1 This practice describes the components of a direct-current-plasma (DCP) optical emission spectrometer. This practice does not attempt to specify component tolerances or performance criteria. This practice does, however, attempt to identify critical factors affecting bias, precision, and sensitivity. A prospective user should consult with the vendor before placing an order to design a testing protocol for demonstrating that the instrument meets all anticipated needs. 1.2 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use. Specific hazards statements are given in Section 9.

Standard Practice for Describing and Specifying a Direct-Current-Plasma Optical Emission Spectrometer

ICS
17.180.30 (Optical measuring instruments)
CCS
N34
发布
1996
实施

この規格は,光波長範囲(波長I00nm一1mm)のレーザ放射バワー及びレーザ放射エネルギーを測定する測定器及び測定装置に適用する。

Power and energy measuring detectors, instruments and equipment for laser radiation

ICS
31.260
CCS
N34
发布
1995-01-01
实施

1.1 This practice is intended as a guide for the use of a flame photometric detector (FPD) as the detection component of a gas chromatographic system.1.2 This practice is directly applicable to an FPD that employs a hydrogen-air flame burner, an optical filter for selective spectral viewing of light emitted by the flame, and a photomultiplier tube for measuring the intensity of light emitted.1.3 This practice describes the most frequent use of the FPD which is as an element-specific detector for compounds containing sulfur (S) or phosphorus (P) atoms. However, nomenclature described in this practice are also applicable to uses of the FPD other than sulfur or phosphorus specific detection.1.4 This practice is intended to describe the operation and performance of the FPD itself independently of the chromatographic column. However, the performance of the detector is described in terms which the analyst can use to predict overall system performance when the detector is coupled to the column and other chromatographic system components.1.5 For general gas chromatographic procedures, Practice E 260 should be followed except where specific changes are recommended herein for use of an FPD.1.6 solely-SI-units;This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use. For specific safety information, see Section , Hazards.

Standard Practice for Using Flame Photometric Detectors in Gas Chromatography

ICS
17.180.30; 71.040.50 (Physicochemical methods of a
CCS
N34
发布
1995
实施

1.1 The purpose of this test method is to obtain data which characterize the sizes of liquid particles or drops such as are produced by a spray nozzle or similar device under specified conditions using a specified liquid. The drops will generally be in the size range from 5-[mu]m to the order of 1000-[mu]m diameter; they will occur in sprays which may be as small as a few cubic centimetres or as large as several cubic metres. Typically the number density of the particles can vary significantly from one point to another. 1.2 This test method is intended primarily for use in standardizing measurements of the performance of spray- producing devices. It is limited to those techniques and instruments that operate by passing a beam of light through the spray and analyzing the light scattered by the droplets to derive size information. Such techniques do not produce images of individual drops, and therefore, are known as "optical (nonimaging) instruments." 1.3 The measurements made, when referred to the entire spray being sampled, may be flux sensitive or spatial, as defined in Practice E799, depending on the techniques used with a particular instrument. 1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

Standard Test Method for Determining Liquid Drop Size Characteristics in a Spray Using Optical Nonimaging Light-Scattering Instruments

ICS
17.180.30 (Optical measuring instruments)
CCS
N34
发布
1995
实施

本标准规定了测长机的产品分类、技术要求、试验方法、检验规则、标志、包装、运输与贮存等。 本标准适用于测量范围为1m、3m和6m的测长机。

Metroscope

ICS
17.180.30
CCS
N34
发布
1994-08-23
实施
1995-05-01

本标准规定了平行平晶的产品分类、技术要求、试验方法、检验规则、标志、包装、运输与贮存等。 本标准适用于以光波干涉方法检验千分尺、杠杆千分尺、杠杆卡规及其他量具测量面的平面度和两测量面的平行度的平行平晶。

Optical parallel

ICS
37.020
CCS
N34
发布
1994-08-23
实施
1995-05-01

本标准规定了平面平晶的产品分类、技术要求、试验方法、检验规则、标志、包装、运输与贮存等。 本标准适用于以光波干涉方法来检验平面的平面度及量块的研合性的圆形平面平晶。

Optical flat

ICS
37.020
CCS
N34
发布
1994-08-23
实施
1995-05-01

本标准规定了光照度计的产品分类、技术要求、试验方法、检验规则、标志、包装、运输、贮存等。 本标准适用于光度头探测器为硅光电池、硅光电二极管和硒光电池的光照度计。光度头探测器为光电管的光照度计亦可参照使用。

Illuminometer

ICS
17.180.30
CCS
N34
发布
1994-08
实施
1995-05-01

Hand saccharimeter

ICS
67.180.10
CCS
N34
发布
1994-07-01
实施



Copyright ©2007-2022 ANTPEDIA, All Rights Reserved
京ICP备07018254号 京公网安备1101085018 电信与信息服务业务经营许可证:京ICP证110310号