11.040.40 外科植入物、假体和矫形 标准查询与下载



共找到 3146 条与 外科植入物、假体和矫形 相关的标准,共 210

规定了增材制造技术制备脊柱侧弯矫形器的基本要求、技术要求、试验方法、包装、运输和交货。

Additive manufacturing spinal orthosis for scoliosis curve-correction

ICS
11.040.40
CCS
C358
发布
2022-09-05
实施
2022-09-05

规定了增材制造制备矫形鞋垫的要求、试验方法、检验规则、标志、包装、运输和贮存。

Additive manufacturing customized orthotic insole

ICS
11.040.40
CCS
C358
发布
2022-09-05
实施
2022-09-05

规定了假肢部件踝足装置储能性的技术要求、试验原则和试验方法。

Requirements and test methods for energy-storing capability of ankle-foot units

ICS
11.040.40
CCS
C358
发布
2022-09-05
实施
2022-09-05

1.1 This guide is intended as a resource for individuals and organizations involved in the development, production, delivery, and regulation of tissue engineered medical products (TEMPs) intended for use in the surgical repair, replacement, shunting, and/or bypass of blood vessels. This guide is intended for use related to the in vitro assessment of TEMP vascular grafts. In vitro cellular characterization and in vivo testing are not within scope for this standard guide. 1.2 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use. 1.3 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

Standard Guide for Characterization and Assessment of Vascular Graft Tissue Engineered Medical Products (TEMPs)

ICS
11.040.40
CCS
发布
2022-09-01
实施

1.1 This practice provides guidelines for short-term testing or screening of candidate materials, both porous and dense, as to the local effects of the material that is implanted intramuscularly. This method may not be applicable for absorbable materials, depending on the absorption profile of the test material. The tissue reactions will be evaluated in comparison to those evoked by control materials that are accepted as clinical implant materials. This is a short-term (less than 30 days) screening procedure for determining acceptability of candidate materials. 1.2 This practice, along with other appropriate biological tests (including other appropriate ASTM tests), may be used in the biocompatibility assessment of the candidate materials for use in the fabrication of devices for clinical application. 1.3 This experimental protocol is not designed to provide a comprehensive assessment of the systemic toxicity, carcinogenicity, or mutagenicity of the material since other standards address these issues. 1.4 This practice is one of several developed for the assessment of the biocompatibility of materials. Practice F748 provides guidance for the selection of appropriate methods for testing materials for a specific application. 1.5 The values stated in SI units, including units officially accepted for use with SI, are to be regarded as standard. No other systems of measurement are included in this standard. 1.6 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use. 1.7 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

Standard Practice for Short-Term Intramuscular Screening of Implantable Medical Device Materials

ICS
11.040.40
CCS
发布
2022-09-01
实施

1.1 This test method specifies the materials and methods for the axial compressive subsidence testing of non-biologic intervertebral body fusion devices, spinal implants designed to promote arthrodesis at a given spinal motion segment. 1.2 This test method is intended to provide a basis for the mechanical comparison among past, present, and future nonbiologic intervertebral body fusion devices. This test method is intended to enable the user to mechanically compare intervertebral body fusion devices and does not purport to provide performance standards for intervertebral body fusion devices. 1.3 This test method describes a static test method by specifying a load type and a specific method of applying this load. This test method is designed to allow for the comparative evaluation of intervertebral body fusion devices. 1.4 Guidelines are established for measuring test block deformation and determining the subsidence of intervertebral body fusion devices. 1.5 Since some intervertebral body fusion devices require the use of additional implants for stabilization, the testing of these types of implants may not be in accordance with the manufacturer’s recommended usage. 1.6 Units—The values stated in SI units are to be regarded as the standard with the exception of angular measurements, which may be reported in terms of either degrees or radians. 1.7 The use of this standard may involve the operation of potentially hazardous equipment. This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use. 1.8 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

Standard Test Method for Measuring Load-Induced Subsidence of Intervertebral Body Fusion Device Under Static Axial Compression

ICS
11.040.40
CCS
发布
2022-09-01
实施

1   Scope This document specifies requirements that are applicable to implantable cardioverter defibrillators and CRT-D s and the functions of active implantable medical devices intended to treat tachyarrhythmia. The tests that are specified in ISO 14708 are type tests and are to be carried out on samples of a device to show compliance. This document was designed for tachyarrhythmia pulse generators used with either endocardial lead s or epicardial lead s. At the time of this edition, the authors recognized the emergence of technologies that do not use endocardial or epicardial lead s for which adaptations of this part will be required. Such adaptations are left to the discretion of manufacturers incorporating these technologies. This document is also applicable to some non-implantable parts and accessories of the devices (see Note 1). The characteristics of the implantable pulse generator or lead shall be determined by either the appropriate method detailed in this document or by any other method demonstrated to have accuracy equal to, or better than, the method specified. In the case of dispute, the method detailed in this document shall apply. Any aspect of an active implantable medical device intended to treat bradyarrhythmias or cardiac resynchronization is covered by ISO 14708‑2. NOTE 1 The device that is commonly referred to as an active implantable med...

Implants for surgery. Active implantable medical devices - Particular requirements for active implantable medical devices intended to treat tachyarrhythmia (including implantable defibrillators)

ICS
11.040.40
CCS
发布
2022-08-31
实施
2022-08-31

1   Scope This document is applicable to active implantable medical devices intended for electrical stimulation of the central or peripheral nervous system. The tests that are specified in this document are type tests and are to be carried out on a sample of a device to assess device behavioural responses, and are not intended to be used for the routine testing of manufactured products.

Implants for surgery. Active implantable medical devices - Implantable neurostimulators

ICS
11.040.40
CCS
发布
2022-08-31
实施
2022-08-31

1   Scope This document specifies particular requirements for active implantable medical devices intended to deliver a medicinal substance to site-specific locations within the human body, to provide basic assurance of safety for both patients and users. It amends and supplements ISO 14708‑1:2014. The requirements of this document take priority over those of ISO 14708‑1. This document is applicable to active implantable medical devices intended to deliver medicinal substances to site-specific locations within the human body. This document is also applicable to some non-implantable parts and accessories of the devices defined in Clause 3. The tests that are specified in this document are type tests intended to be carried out on a sample of a device to show compliance and are not intended to be used for the routine testing of manufactured products. NOTE This document is not intended to apply to non-implantable infusion systems.

Implants for surgery. Active implantable medical devices. - Implantable infusion pump systems

ICS
11.040.40
CCS
发布
2022-08-31
实施
2022-08-31

1   Scope This document specifies requirements that are applicable to those active implantable medical devices intended to treat bradyarrhythmias and devices that provide therapies for cardiac resynchronization. The tests that are specified in this document are type tests, and are to be carried out on samples of a device to show compliance. This document was designed for bradyarrhythmia pulse generators used with endocardial leads or epicardial leads . At the time of this edition, the authors recognized the emergence of leadless technologies for which adaptations of this part will be required. Such adaptations are left to the discretion of manufacturers incorporating these technologies. This document is also applicable to some non-implantable parts and accessories of the devices (see Note 1). The electrical characteristics of the implantable pulse generator or lead are determined either by the appropriate method detailed in this particular standard or by any other method demonstrated to have an accuracy equal to, or better than, the method specified. In case of dispute, the method detailed in this particular standard applies. Any features of an active implantable medical device intended to treat tachyarrhythmias are covered by ISO 14708‑6. NOTE 1 The device that is commonly referred to as an active implantable medical device can in fact be a single device, a combination of devices, or a combinat...

Implants for surgery. Active implantable medical devices - Cardiac pacemakers

ICS
11.040.40
CCS
发布
2022-08-31
实施
2022-08-31

1   Scope This document specifies requirements for safety and performance of active implantable circulatory support devices , including type tests, animal studies and clinical evaluation requirements. NOTE The device that is commonly referred to as an active implantable medical device can in fact be a single device, a combination of devices, or a combination of a device or devices and one or more accessories. Not all of these parts are required to be either partially or totally implantable, but there is a need to specify main requirements of non-implantable parts and accessories if they could affect the safety or performance of the implantable device. The tests that are specified in this document are type tests and are to be carried out on a sample of a device to assess device behavioural responses and are not intended to be used for the routine testing of manufactured products. Included in the scope of this document are: — ventricular assist devices ( VAD ), left or right heart support; — total artificial hearts ( TAH ); — biventricular assist devices ( b...

Implants for surgery. Active implantable medical devices - Circulatory support devices

ICS
11.040.40
CCS
发布
2022-08-31
实施
2022-08-31

1   Scope This document specifies requirements that are applicable to those active implantable medical devices that are intended to treat hearing impairment via electrical stimulation of the auditory pathways. Devices which treat hearing impairment via means other than electrical stimulation are not covered by this document. The tests that are specified in this document are type tests and are to be carried out on samples of a device to show compliance. This document is also applicable to non-implantable parts and accessories of the devices (see NOTE). The electrical characteristics of the implantable part are determined by either the appropriate method detailed in this document or by any other method demonstrated to have an accuracy equal to, or better than, the method specified. In the case of dispute, the method detailed in this document applies. NOTE A device that is commonly referred to as an active implantable medical device can in fact be a single device, a combination of devices, or a combination of a device or devices and one or more accessories. Not all of these parts are required to be either partially or totally implantable, this document specifies those requirements of non-implantable parts and accessories which could affect the safety or performance of the implantable part.

Implants for surgery. Active implantable medical devices - Particular requirements for cochlear and auditory brainstem implant systems

ICS
11.040.40
CCS
发布
2022-08-31
实施
2022-08-31

本文件规定了人工关节用超高分子量聚乙烯(PE-UHMW)树脂的分类、性能要求、试验方法、检验规则及标志、包装、运输和贮存。 本文件适用于人工关节用超高分子量聚乙烯(PE-UHMW)树脂。

Ultra-high-molecular-weight polyethylene (PE-UHMW) resin for joint replacements

ICS
11.040.40
CCS
C292
发布
2022-08-30
实施
2022-09-02

本文件规定了外科植入物用大剂量辐照交联超高分子量聚乙烯(以下简称交联超高分子量聚乙烯)制品的要求及试验方法。 本文件涉及满足GB/T 19701.1、GB/T 19701.2所有要求的超高分子量聚乙烯树脂粉料、加工成型的模塑料(棒材、板材),以及交联制品。 本文件仅适用于经过40 kGy及以上剂量的伽马射线或电子束辐照的超高分子量聚乙烯制品。

Standard guide for extensively irradiation-crosslinked ultra-high molecular weight polyethylene fabricated form for surgical implant applications

ICS
11.040.40
CCS
C358
发布
2022-08-04
实施
2022-08-04

This document specifies requirements that are applicable to those active implantable medical devices that are intended to treat hearing impairment via electrical stimulation of the auditory pathways. Devices which treat hearing impairment via means other than electrical stimulation are not covered by this document. The tests that are specified in this document are type tests and are to be carried out on samples of a device to show compliance. This document is also applicable to non-implantable parts and accessories of the devices (see NOTE). The electrical characteristics of the implantable part are determined by either the appropriate method detailed in this document or by any other method demonstrated to have an accuracy equal to, or better than, the method specified. In the case of dispute, the method detailed in this document applies. NOTE A device that is commonly referred to as an active implantable medical device can in fact be a single device, a combination of devices, or a combination of a device or devices and one or more accessories. Not all of these parts are required to be either partially or totally implantable, this document specifies those requirements of non-implantable parts and accessories which could affect the safety or performance of the implantable part.

Implants for surgery - Active implantable medical devices - Part 7: Particular requirements for cochlear and auditory brainstem implant systems (ISO 14708-7:2019)

ICS
11.040.40
CCS
发布
2022-07-27
实施
2022-10-31

This document specifies requirements that are applicable to implantable cardioverter defibrillators and CRT-Ds and the functions of active implantable medical devices intended to treat tachyarrhythmia. The tests that are specified in ISO 14708 are type tests and are to be carried out on samples of a device to show compliance. This document was designed for tachyarrhythmia pulse generators used with either endocardial leads or epicardial leads. At the time of this edition, the authors recognized the emergence of technologies that do not use endocardial or epicardial leads for which adaptations of this part will be required. Such adaptations are left to the discretion of manufacturers incorporating these technologies. This document is also applicable to some non-implantable parts and accessories of the devices (see Note 1). The characteristics of the implantable pulse generator or lead shall be determined by either the appropriate method detailed in this document or by any other method demonstrated to have accuracy equal to, or better than, the method specified. In the case of dispute, the method detailed in this document shall apply. Any aspect of an active implantable medical device intended to treat bradyarrhythmias or cardiac resynchronization is covered by ISO 14708-2. NOTE 1 The device that is commonly referred to as an active implantable medical device can in fact be a single device, a combination of devices, or a combination of a device or devices and one or more accessories. Not all of these parts are required to be either partially or totally implantable, but there is a need to specify some requirements of non-implantable parts and accessories if they could affect the safety or performance of the implantable device. NOTE 2 In this document, terms printed in italics are used as defined in Clause 3. Where a defined term is used as a qualifier in another term, it is not printed in italics unless the concept thus qualified is also defined.

Implants for surgery - Active implantable medical devices - Part 6: Particular requirements for active implantable medical devices intended to treat tachyarrhythmia (including implantable defibrillators) (ISO 14708-6:2019)

ICS
11.040.40
CCS
发布
2022-07-27
实施
2022-10-31

This document specifies particular requirements for active implantable medical devices intended to deliver a medicinal substance to site-specific locations within the human body, to provide basic assurance of safety for both patients and users. It amends and supplements ISO 14708-1:2014. The requirements of this document take priority over those of ISO 14708-1. This document is applicable to active implantable medical devices intended to deliver medicinal substances to site-specific locations within the human body. This document is also applicable to some non-implantable parts and accessories of the devices defined in Clause 3. The tests that are specified in this document are type tests intended to be carried out on a sample of a device to show compliance and are not intended to be used for the routine testing of manufactured products. NOTE This document is not intended to apply to non-implantable infusion systems.

Implants for surgery - Active implantable medical devices - Part 4: Implantable infusion pump systems (ISO 14708-4:2022)

ICS
11.040.40
CCS
发布
2022-07-27
实施
2022-10-31

ISO 14708-3:2017 is applicable to active implantable medical devices intended for electrical stimulation of the central or peripheral nervous system. The tests that are specified in this document are type tests and are to be carried out on a sample of a device to assess device behavioural responses, and are not intended to be used for the routine testing of manufactured products.

Implants for surgery - Active implantable medical devices - Part 3: Implantable neurostimulators (ISO 14708-3:2017)

ICS
11.040.40
CCS
发布
2022-07-27
实施
2022-10-31

This document specifies requirements that are applicable to those active implantable medical devices intended to treat bradyarrhythmias and devices that provide therapies for cardiac resynchronization. The tests that are specified in this document are type tests, and are to be carried out on samples of a device to show compliance. This document was designed for bradyarrhythmia pulse generators used with endocardial leads or epicardial leads. At the time of this edition, the authors recognized the emergence of leadless technologies for which adaptations of this part will be required. Such adaptations are left to the discretion of manufacturers incorporating these technologies. This document is also applicable to some non-implantable parts and accessories of the devices (see Note 1). The electrical characteristics of the implantable pulse generator or lead are determined either by the appropriate method detailed in this particular standard or by any other method demonstrated to have an accuracy equal to, or better than, the method specified. In case of dispute, the method detailed in this particular standard applies. Any features of an active implantable medical device intended to treat tachyarrhythmias are covered by ISO 14708-6. NOTE 1 The device that is commonly referred to as an active implantable medical device can in fact be a single device, a combination of devices, or a combination of a device or devices and one or more accessories. Not all of these parts are required to be either partially or totally implantable, but there is a need to specify some requirements of non-implantable parts and accessories if they could affect the safety or performance of the implantable device. NOTE 2 In this document, terms printed in italics are used as defined in Clause 3. Where a defined term is used as a qualifier in another term, it is not printed in italics unless the concept thus qualified is also defined.

Implants for surgery - Active implantable medical devices - Part 2: Cardiac pacemakers (ISO 14708-2:2019)

ICS
11.040.40
CCS
发布
2022-07-27
实施
2022-10-31

This document specifies requirements for safety and performance of active implantable circulatory support devices, including type tests, animal studies and clinical evaluation requirements. NOTE The device that is commonly referred to as an active implantable medical device can in fact be a single device, a combination of devices, or a combination of a device or devices and one or more accessories. Not all of these parts are required to be either partially or totally implantable, but there is a need to specify main requirements of non-implantable parts and accessories if they could affect the safety or performance of the implantable device. The tests that are specified in this document are type tests and are to be carried out on a sample of a device to assess device behavioural responses and are not intended to be used for the routine testing of manufactured products. Included in the scope of this document are: — ventricular assist devices (VAD), left or right heart support; — total artificial hearts (TAH); — biventricular assist devices (biVAD); — percutaneous assist devices; — paediatric assist devices.

Implants for surgery - Active implantable medical devices - Part 5: Circulatory support devices (ISO 14708-5:2020)

ICS
11.040.40
CCS
发布
2022-07-27
实施
2022-10-31



Copyright ©2007-2022 ANTPEDIA, All Rights Reserved
京ICP备07018254号 京公网安备1101085018 电信与信息服务业务经营许可证:京ICP证110310号